10 Specifications
86500-IMG rev 3
GE Healthcare
71
10 Specifications
System specifications for Diamond, model MGX 2000
Manufacturer
GE Healthcare Finland Oy
Kuortaneenkatu 2, FIN-00510
Helsinki, Finland
Manufacturer's quality system
according to
ISO9001
Manufacturer's environmental
management system according
to
ISO14001
General
Medical electrical equipment
Model
MGX 2000
Power supply
Mains plug connection, 1-phase
(line, neutral and protective
ground, or L1, L2 and protective
ground)
Apparent resistance of supply
mains
0.75
Ω
Rated voltage
230 VAC ±10%, single phase, 50/
60 Hz
16 A slow blow fuse in power
outlet
NOTE!
The Diamond is a cord-connected device, delivered without a plug (in
some markets). Depending upon local electrical code requirements, it
may be either hard-wired or plugged into an outlet. Please consult your
electrician to determine the requirements, and if necessary to provide a
plug which matches your outlet.
Line Regulation:
± 2% (full load)
Branch Circuit:
dedicated run 20A, 2-pole circuit
breaker L1/L2 (10 AWG)
Outlet per local electrical code
requirements or NEMA L6-20
Grounding:
0V potential (earth), continuous
wire run (not conduit)
Rated current
20 A Momentary (during exposure)
300 mA long term (otherwise)
Rated input power
4.6 kVA
IEC safety class
I
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).