5 Operating procedures
86500-IMG rev 3
GE Healthcare
31
5.7.1
ECS movement
1
The C-Arm is driven upwards without applying compression, to open
the infra-mammary fold.
2
The breast is compressed slightly with the paddle to hold it in place.
3
Final compression is applied by the Bucky or cassette tunnel from
below using the ECS compression movement.
5.7.2
Cranio-Caudal (CC) Projection
1
Select the required film size, load the film into the cassette, and place
the cassette into the cassette holder.
2
Make sure that the C-Arm is in the upright position (0º) by pressing
the CC-key. Adjust the image receptor height using the vertical
height control so that patient’s nipple is in profile, no skin folds are
present under the breast and that the infra-mammary fold is open.
Refer to Fig. 5.8, Correct vertical position and Fig. 5.9, Incorrect
Fig 5.8.
Correct vertical position
Fig 5.9.
Incorrect vertical position