8 Troubleshooting
62
GE Healthcare
86500-IMG rev 3
Er0059
Anode run failure.
Clear and try again. If
problem persists, call service.
Er0060
Exposure button
released during
exposure.
Load new film and expose.
Hold exposure button down
until the audio signal ends.
Er0062
Generator failure. kV not
OK.
Clear and try again. If
problem persists, call service.
Er0063
Line voltage low.
Have an electrician check the
voltage of the electric outlet.
If it is strong and stable
enough, then call service for
further actions.
Er0064
Cassette missing.
Insert a cassette and try also
with another type.
Er0065
Exposure time exceeds
limit (too long/too short).
Increase/decrease kV values
and try again. If problem
persists, call service.
Er0066
Movement button or
switch exposure button
pressed during
movement or start-up,
function not allowed.
Clear and try again. If
problem persists, call service.
Er0067
Device not ready for
exposure.
Clear and try again. If
problem persists, call service.
Er0068
Delta 32 PC refused to
receive exposure.
Clear and try again. If
problem persists, call service.
Er0069
Delta 32 PC connection
time-out.
Clear and try again. If
problem persists, call service.
Er0070
Error in stereo motor
pulse measuring.
Clear and try again. If
problem persists, call service.
Er0071
Tube head heat limit
exceeded.
Wait and let cool down
enough. Remaining cooling
time indicated in timer screen.
Clear and try again. If
problem persists, call service.
Er0072
Side collimator failure.
Clear and try again. If
problem persists, call service.
Er0074
Internal communication
error. AEC Board not
answering. Cable
broken.
Clear and try again. If
problem persists, call service.
Error code
in LCD
display
Reason for error
Corrective action
Table 8.1 Diamond error code guide