2
CARESCAPE PDM
Cleaning and disinfection overview
The following cleaning and disinfection information applies to devices, device
components, supplies, and accessories manufactured by GE.
The information provided in this chapter does not supersede any instructions for use
provided by the manufacturer or provided with a device, device component, supply,
or accessory.
For cleaning, disinfection, and care information for devices, device components,
supplies, and accessories made by manufacturers other than GE, see the applicable
instructions for use provided by the manufacturer.
Cleaning
Cleaning warnings
WARNING
Before cleaning or disinfecting, disconnect the device from
the power source.
WARNING
Avoid using chemicals other than the ones described in this
document, as they may damage device surfaces or labels,
or cause equipment failures.
WARNING
Do not pour or spray any liquid directly on the device, cables,
or leadwires or permit fluid to seep into connections or
openings.
WARNING
Never immerse any part of the device, cables, or leadwires in
liquids or allow liquid to enter the interior of the device.
WARNING
Do not autoclave any part of the system with steam (including
cables or leadwires) or sterilize with ethylene oxide.
WARNING
Never use conductive solutions, oxidizing compounds, wax, or
wax compounds to clean device.
WARNING
If liquid has accidentally entered the system or its parts,
disconnect the power cord from the power supply, remove
the battery, and have the equipment serviced by qualified
service personnel.
2110145-001
Cleaning and disinfecting
11
Summary of Contents for CARESCAPE B105
Page 8: ...8 Cleaning and disinfecting 2110145 001 ...
Page 30: ...ApexPro and CARESCAPE Telemetry 30 Cleaning and disinfecting 2110145 001 ...
Page 48: ...CARESCAPE B450 Monitor 48 Cleaning and disinfecting 2110145 001 ...
Page 54: ...CARESCAPE B650 Monitor 54 Cleaning and disinfecting 2110145 001 ...
Page 71: ...content ...