ApexPro and CARESCAPE Telemetry
●
Never connect any device or applied part to a patient until it is thoroughly dry.
●
If you discover any signs of deterioration or damage in the device, discontinue its
use.
General cleaning instructions
Follow these instructions to clean the device and other non-applied parts unless the
other parts provide part-specific instructions. Non-applied parts refer to those parts of
the system that are not in direct contact with the patient.
Telemetry monitoring devices
1.
Remove all cables and the batteries.
2.
Close the battery door.
3.
Dampen a soft lint-free cloth with one of the permitted detergents.
4.
Wring excess liquid from the cloth and wipe the exterior surface. Pay special
attention to hard-to-clean areas, like grooves and crevices.
●
Remove any soil as soon as possible after use by wiping the device.
●
Any contact of cleaning solutions with metal parts may cause corrosion.
●
Do not damage or bend connector pins when cleaning or drying.
●
Do not let fluid pool around connection pins. If this happens, blot dry with
a cotton swab or soft cloth.
5.
Wipe off the cleaning solutions with a clean, lightly moistened cloth.
6.
Visually ensure that the device is clean. If visible soil remains, repeat cleaning
procedure until the device is thoroughly clean.
7.
Dry thoroughly with a dry, lint-free cloth and let air dry for at least 30 minutes.
Drying times may vary based on the environmental conditions.
8.
Install the batteries and close the battery door.
9.
Reconnect all cables.
Battery compartment cleaning (ApexPro FH transceivers only)
NOTICE
The battery compartment is not waterproof.
NOTICE
Make sure fluids do not enter the electronics through the air
holes in the battery compartment floor.
NOTICE
Cleaning the battery compartment in a manner other than
that specified may cause the unit to malfunction and void
the warranty.
Under normal conditions, the battery compartment should not require cleaning. If the
battery compartment does require cleaning, use the following instructions.
1.
Remove the battery from the battery compartment.
2.
Dampen a soft lint-free cloth with one of the permitted detergents.
3.
Wring excess liquid from the cloth and wipe the interior of the battery
compartment. Make sure moisture does not enter the electronics area below
the battery compartment floor.
2110145-001
Cleaning and disinfecting
25
Summary of Contents for CARESCAPE B105
Page 8: ...8 Cleaning and disinfecting 2110145 001 ...
Page 30: ...ApexPro and CARESCAPE Telemetry 30 Cleaning and disinfecting 2110145 001 ...
Page 48: ...CARESCAPE B450 Monitor 48 Cleaning and disinfecting 2110145 001 ...
Page 54: ...CARESCAPE B650 Monitor 54 Cleaning and disinfecting 2110145 001 ...
Page 71: ...content ...