Figure 5.7: Observing the bed height straight on rather from at an angle.
f
When the compression is completed the system automatically stops the flow of
packing liquid.
g
Read the final bed height according to the scale on the side of the column.
There are different scales for stainless steel and plastic bed supports
on the column.
Note:
5
When the packing test is completed open the Evaluation module of UNICORN and
perform an integration of the peak using the packed bed height value.
6
It is recommended to replace the tubing with a stop plug [101] at the check valve
[116].
78
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
5 Operation
5.1 Intelligent Packing of the column with ÄKTAexplorer and ÄKTApilot