Tubing used between column
and system
Table 4.2: Tubing internal diameters for connection between column and ÄKTA
system.
Tubing i.d. (mm)
ÄKTA system
Column diameter (mm)
1.2
explorer
50
1.7
pilot
50
1.7
avant
50
1.7
avant
70
1.7, 2.9
pilot
70
2.9
pilot
100
4.10
Installing the UNICORN Intelligent Packing strategy
on ÄKTAexplorer and ÄKTApilot
Install the correct UNICORN strategy containing Intelligent Packing for the appropriate
system according to the instructions on the cover of the strategy CD supplied with the
column. The latest Wizard can be downloaded from the AxiChrom web page, see Sec-
tion 8.2 Contact information, on page 155.
60
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
4 Getting started
4.9 Installation in basic configuration