Certification & Safety Statements for GE Measurement & Control Ultrasonic Flow Transmitters
When installing this apparatus, the following requirements must be met:
• Field wiring shall be rated at least 10°C above 70°C.
• Connecting cables shall be mounted securely and protected from mechanical damage, pulling and twisting.
• Cable entries are ¾” NPT.
• Cable glands of an approved flameproof design are required. These must be installed according to the manufacturer’s
instructions. Where the cable glands are provided by GE, the manufacturer’s instructions, as supplied, to GE, will be
included in the documentation.
• Unused cable entries must be sealed using a certified threaded plug.
• Modifications to the flameproof enclosure are not permitted.
• The apparatus should be de-energized before opening.
• Installation should comply with IEC/EN 60079-14.
• Equipment is of type flameproof “d” design and complies with: EN 60079-0:2009, EN 60079-1:2007,
EN 60529:1991 +A1:2000, IEC 60079-0:2011, IEC 60079-1:2007, IEC 60529:2001.
• The product contains no exposed parts which produce surface temperature infrared, electromagnetic ionizing, or
non-electrical dangers.
• The product must not be subjected to mechanical or thermal stresses in excess of those permitted in the certification
documentation and the instruction manual.
• The product cannot be repaired by the user; it must be replaced by an equivalent certified product. Repairs should only
be carried out by the manufacturer or by an approved repairer.
• Only trained, competent personnel may install, operate and maintain the equipment
• The product is an electrical apparatus and must be installed in the hazardous area in accordance with the requirements
of the EC Type Examination Certificate. The installation must be carried out in accordance with all the appropriate
international, national and local standard codes and practices and site regulations for flameproof apparatus and in
accordance with the instructions contained in the manual. Access to the circuitry must not be made during operation.
Special Conditions for Safe Use:
Consult the manufacturer if dimensional information on the flameproof joints is
necessary.
Markings:
Markings shall appear on the product as shown below:
WATTS: XXX
VOLTAGE: XXX
US
APPROVED
MFG DATE: MMXXYY
S/N: XXXXXXXX
MODEL: XXXXXXX
C
FM
Tamb = -20°C to +60°C
FM12ATEX0040X
II 2 G Ex d IIB + H2 T6 Gb IP66
AT LEAST 10°C ABOVE 70°C.
SUPPLY CONNECTION WIRING SHALL BE RATED
CABLE GLAND ENTRIES ARE 3/4" NPT
CL I, DIV. 1, GPS BCD; T6.
IECEx FMG12.0012X
BILLERICA MA USA
1180
GE INFRASTRUCTURE SENSING
CL I, ZONE 1, AEx/Exd IIB + H2
1100 TECHNOLOGY PARK DR.
WARNING - DO NOT OPEN WHEN ENERGIZED
AVERTISSEMENT - Ne pas ouvrir sous tension
WARNING - DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE IS PRESENT
AVERTISSEMENT - Ne pas ouvrir une atmosphere eplosive gazeuse est presente
Summary of Contents for AquaTrans AT600
Page 1: ...Flow 910 312 Rev D September 2017 bhge com AquaTrans AT600 User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x AquaTrans AT600 User s Manual no content intended for this page ...
Page 34: ...Chapter 2 Clamp On Fixture and Transducer Installation 22 AquaTrans AT600 User s Manual ...
Page 107: ...AquaTrans AT600 User s Manual 95 Chapter 5 Error Codes and Troubleshooting ...
Page 108: ...Chapter 5 Error Codes and Troubleshooting 96 AquaTrans AT600 User s Manual ...
Page 183: ...AquaTrans AT600 User s Manual 171 Chapter 6 Communication ...
Page 184: ...Chapter 6 Communication 172 AquaTrans AT600 User s Manual ...
Page 190: ...Appendix A Specifications 178 AquaTrans AT600 User s Manual no content intended for this page ...
Page 194: ...Appendix B Data Records 182 AquaTrans AT600 User s Manual no content intended for this page ...
Page 212: ... no content intended for this page ...
Page 214: ... no content intended for this page ...
Page 215: ......