ULTRATHERM 908i
24
gbo Medizintechnik AG 2004
Version 1.2
•
Measure the voltage drops between the earthing conductor of the mains connector and all metal
parts of the unit, that can be touched, under the following condition:
I
earth
= 10 A, U
earth/setpoint
< 2 V.
•
Record the maximum measurement value.
•
Perform the earthing conductor test according to DIN 57 751 or VDE 0751, respectively.
•
Record the measurement values.
4.11.4. Measurement of the mains current consumption
•
Measure the mains current consumption according to clause 4.6 and report the value.
4.11.5. Check of the power values
•
Measure the power values according to clauses 4.8.1. and 4.8.2. and report the values.
4.11.6. Operation check
•
Perform the operation check according to clause 4.4.