HiToP 1touch

 

 gbo Medizintechnik

 

AG

 

Version 1.2 

4

 

Therapy 

4.1

 

General guidelines for therapy 

The

 

HiToP

 

1touch

 

guides the user through the application of the therapy. With the 

direct help menu all buttons and their functions are explained directly on the screen. 
 
On the right side of the screen therapy programs can be selected directly from this list 
of favorites or with the sub-menus “More Therapies” or “Indications”. 
 
Select  the  desired  program  or  indication.  The  direct  help  menu  will  give  you 
information about the following steps of the therapy.  
 
The HiToP

®

-therapy is a special treatment and it is very different from the classical 

electrotherapy. In the HiToP

®

-therapy, the application technique for the electrodes is 

more  complicated  and,  typically,  the  treatment  time  is  longer  than  in  the  classical 
electrotherapy.  We  talk  about  a  holistic  treatment.  For  this  reason,  the  treatment 
suggestions  in  the  list  of  indications  are  mostly  structured  in  two  steps.  In  the  first 
step or main treatment, the treatment area is, in most cases, flooded with SimulFAM i 
transversally. In the Follow-Up therapy, treatment is often a whole body treatment or 
it  is  the  treatment  with  SimulFAM  X  longitudinally.  With  this  method,  the 
therapeutic  success  which  is  accomplished  by  multi-circuit  treatment  using  a  multi-
channel device (2 or 4-channel) can be achieved by providing a sequential treatment 
with the single channel HiToP

®

 1Touch. 

Typically,  you  have  to  change  the  electrode  placement  after  the  first  half  of  the 
treatment in order to start the Follow-Up treatment. 
 
If such a treatment is not possible in the daily practice routine, the treatment can be 
switched to a single step treatment. In this case, only the main treatment step is used 
for the therapy. This therapy lasts as long as both steps of the Follow-Up therapy. 
 
 
The time period of the treatment for the indications can be selected in the settings. 
 
 

 

4.2

 

Selection of therapies 

4.2.1

 

Selection of therapy by More Therapies 

In this selection all pre-defined and all user-defined programs are filed. The selection 
is made from a list which is filtered. Thus, only pre-defined or user-defined currents 
can be displayed from the appropriate list and then selected. 

 

Summary of Contents for HiToP 1touch

Page 1: ...HighTone Therapy Device HiToP 1touch Instructions for use...

Page 2: ...rrors or omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of gbo Medizintechni...

Page 3: ...on 13 3 1 5 Screen for selecting More Therapies Standard Therapies 13 3 1 6 Screen for selecting More Therapies Own Therapies 14 3 2 Direct help 15 3 3 Settings 16 3 4 Intensity monitoring 18 3 4 1 De...

Page 4: ...S AND SAFETY PRECAUTIONS 29 7 1 Contraindications 30 8 EXPLANATION OF THE PICTOGRAPHS USED 31 9 TECHNICAL DATA 32 9 1 Current Types 33 10 ACCESSORIES 34 11 TROUBLESHOUTING 35 12 APPENDIX A STANDARD PR...

Page 5: ...eatment for overall vitalisation muscle relaxation muscle training muscle stimulation increase of energy consumption rehabilitation Also patients with varicose veins metal implants endoprothesis and o...

Page 6: ...the respective socket on the back side of the device 3 The display shows the window with the settings of the last treatment 4 Select SimulFAM i from the list of More Therapies or the favorites list 5...

Page 7: ...e current stimulation frequency is shown The frequency values increase and decrease every second 8 A special tone marks the end of the treatment time 9 The current to the patient will decrease automat...

Page 8: ...nsport use always the designated HiToP 1touch transport bag In this bag enough space for all the accessories is available Also for the Aloe Vera contact spray a bottle holder is provided 2 1 3 General...

Page 9: ...designed to be connected to a supply voltage of 100 to 240 V It is not necessary to switch over the voltage the device adjusts automatically to the right voltage Irrespective of the adjusted supply vo...

Page 10: ...h is equipped with a SD card slot located to the left of the display 2 3 Placing out of order In order to disconnect the device just disconnect it from the mains power supply Before the disposal of th...

Page 11: ...ration Consequently operation is explained in broad outline only 3 1 Control Panel 3 1 1 Touch screen The touch screen should be operated with a distinct pressure of the fingertip Pointed objects are...

Page 12: ...button such as the changeable frequency When the button is pressed it is provided with a blue border and the adjusting knob may be operated for 3 seconds 3 1 3 Home screen After power up and ready for...

Page 13: ...lectrode placement Indications list in alphabetical order Selection of treatment area to restrict the list Return to the Home screen without accepting the indication Accept the selected indication and...

Page 14: ...n Therapies the device displays the following screen List of own therapies in alphabetical order Parameters for the selected list entry Return to the Home screen without accepting the own therapy Acce...

Page 15: ...lp function first touch the button and then the button for which you desire information A window opens where the information is shown Touch to the help window in order to close it Note Available butto...

Page 16: ...no therapy is active you can switch to the settings menu The following items are available in the Settings menu Display Setting the national language Calibration of the touch screen Audio Setting the...

Page 17: ...search for faults in the hardware Note The basic settings for the duration of SimulFAM i SimulFAM X and the Standard Therapies may be changed in the selection menu of More Therapies Settings Possible...

Page 18: ...ensity anew Abrupt resistance changes in the electrode application could cause problems 1 Two electrodes touch each other and form a shunt during treatment If the patient moves during the therapy that...

Page 19: ...ensity monitoring as an exception e g in case of the Big Tast Therapy 1 Press the button Intensity monitoring 2 The blue bar which indicates Intensity monitoring is active will get grey Note If the In...

Page 20: ...reatment area is in most cases flooded with SimulFAM i transversally In the Follow Up therapy treatment is often a whole body treatment or it is the treatment with SimulFAM X longitudinally With this...

Page 21: ...set the parameters individually This therapy can now be stored by pressing and holding 2 5 seconds the corresponding button on the right location of the favorites list Entries which are no longer requ...

Page 22: ...be stopped and released upon new activation The frequency set may be changed with the bar beneath the graphics The intensity of SimulFAM i may also be changed during therapy 4 4 Setting SimulFAM X If...

Page 23: ...ectrode 50 mm x 104 mm 52 cm2 104 mA Table 1 Maximum currents for electrodes 5 1 General notes on conductive rubber electrodes We recommend the use of conductive rubber electrodes in various sizes For...

Page 24: ...o high or the electrodes were used for too many treatments old and worn electrodes Consequently it might be possible that the adhesive electrodes get warm during the therapy and even heat up to a temp...

Page 25: ...a way that the electrode covers the skin all over 4 Connect the electrodes to the patient cable 5 Connect the patient cable to the distribution cable 6 Plug in the patient plug of the distribution ca...

Page 26: ...vice are allowed to be serviced while a patient is connected 6 1 Safety Controls The device is subject to the provisions of the Medical Device Directive The safety controls have to be carried out on t...

Page 27: ...gases The HiToP 1touch is suited for wiping disinfection Make sure that no liquids soak into the device Under no circumstances the plug or socket must get wet Do not sprinkle the device for cleaning o...

Page 28: ...ck side of the electrodes leave to soak and wipe off with a lint free cloth Please follow the recommendations of the manufacturer We recommend Bacillol AF for the disinfection of the electrodes Note T...

Page 29: ...he patient cable is in the direct range of a high frequency short wave or micro wave therapeutic device damage to the device or an injury to the patient cannot be excluded Please keep a clearance of 3...

Page 30: ...pacemaker the treatment at the lower extremities is not contraindicated Note Patients with organ transplantation who take immunesuppresive medications should not be treated with SimulFAM i Note The a...

Page 31: ...r use Follow operating instructions Caution Application part ungrounded protection degree Type BF This product complies with WEEE Directive 2002 96 EG waste electrical and electronic equipment Separat...

Page 32: ...cm x 6 cm W x H x D Weight 2 44 kg without accessories Color aluminium natural anodized and grey RAL 7016 Display 10 1 TFT LCD touch screen Environmental conditions operation of the device temperatur...

Page 33: ...described alternating currents are modulated by frequencies in the range of 0 1 200 Hz The output current is free of dc parts The maximum output voltage is 76 V and the maximum current is 310 mA The m...

Page 34: ...red 6 cables of 100 cm each 017 0 0034 Conductive rubber electrode 115 x 175 package of 2 017 0 0046 Conductive rubber electrode 80 x 120 package of 2 017 0 0047 Elastic straps with Velcro 70 cm packa...

Page 35: ...thorized by the manufacturer Further errors symptom cause and action The device cannot be switched on no display is shown Please check the mains plugs and sockets If necessary contact your service age...

Page 36: ...Muscle Stimulation HTEMS 20 Hz 3s 3s 3s Changeable in More Therapies Synchr Stim 20 Hz 3s 3s 7s Changeable in More Therapies Lipolysis 2 5 3 5 Hz Changeable in More Therapies Cellulite 2 5 3 5 Hz 10 m...

Page 37: ...manufacturer projects a product life of 10 years HiToP 1touch is an electronic device Waste disposal is to be done according to the regulations for electronic devices Consumables have to be disposed...

Page 38: ...test IEC 60601 test level Compliance level Electromagnetic environment guidance electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be made of wood concr...

Page 39: ...or 80 MHz to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation...

Page 40: ...Electrode positioning 25 Electrodes 23 27 H Home screen 12 I Indications 13 Intended Use 5 Intensity and modification knob 12 M Maintenance 26 O Operating 11 S Safety controls 26 SD card 10 Settings...

Page 41: ...HiToP 1touch 41 gbo Medizintechnik AG Version 1 2 Notes...

Page 42: ...42 HiToP 1touch gbo Medizintechnik AG Version 1 2 Notes...