HiToP
®
191
15
gbo Medizintechnik AG
2005
Version 1.5
5
Electrodes
The standard accessories for HiToP
191 include rubber electrodes of 120 x 80
mm.
5.1
Electrode application
The positioning of the patient is very important. The therapy shall always be
carried out in a comfortable and relaxed position. The joints shall be placed in an
angular position, so that they are placed in a relaxed position, from which both the
flexor as well as the extensor muscles can be stimulated.
In general, the preparation and application of the electrodes is carried out as
follows:
1. Connect the electrodes to the patient cable.
2. Connect the patient cable to the distribution cable.
3. Plug in the patient plug of the distribution cable to the socket of the unit.
4. Wet the conductive rubber electrodes with Aloe contact spray to avoid current
sensations on the skin.
5. Remove oil and creme of the skin with a cleaning tissue.
6. Apply the electrodes to the patient.
7. Apply the Velcro strap in a way that the electrode fits completely.
Attention
•
Do not apply the electrodes on skin injuries. Even minor abrasions can
cause a burning to the patient. Also, the patient may not be able to
judge the current intensity objectively. If this cannot be avoided, apply
zinc ointment or vaseline to cover the affected parts of the skin.
•
The electrodes must adhere completely to the skin in order to avoid
excessive local current density. Otherwise there can occur erythemas,
and sometimes even burns.