12
HiToP
®
191
gbo Medizintechnik AG
2005
Version 1.5
4
Therapy
4.1
General notes about the therapy
For the therapy we use a muscle stimulation. The muscle stimulation is generated
with HiToP
frequencies. The current is applied by rubber electrodes which are
moistured with aloe vera contact spray.During the treatment the muscle is
contracted in intervals. An interval consists of a 3 seconds rise time (intensity
increases), a 3 seconds hold time (intensity is constant on the maximum setted
before) and a 3 seconds break time (Intensity = 0).
!!
Note
If you like to interrupt or stop the treatment, decrease the
intensity to 0 by turning the intensity knob counterclockwise.
4.2
Treatment details
1.
Switch on the unit
2.
You see the start display:
3.
In the display the following message appears:
Adjust the intensity
until you get a
muscle contraction.
4.
Increase the intensity clockwise until you see a visible muscle movement.
5.
In the display the following message appears:
TIME: 59:22
INT: 165 mA*
6.
The therapy starts. The therapy timer is shown in the display and decreases in
steps of seconds.
7.
With the white softkey-button you can select between the the overview display
and the detailed view.