Guidance and manufacturer’s declaration — electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment should assure that it is used in such an environment.
Immunity test
IEC 60601- test level Compliance level
Electromagnetic environment –
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the equipment, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
Conducted RF,
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
3 V
eff
d=1,2
√
P
Radiated RF,
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3V/m
d=1,2
√
P
for 80 MHz to 800 MHz
d=2,3
√
P
for 800 MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts according
to the transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Recommended separation distances to portable and mobile RF communication equipment
The equipment is intended to be operated in an electromagnetic environment, where radiated RF
interference is controlled. The user can help in avoiding interferences by means of meeting minimum
separation distances between portable and mobile RF communication equipment (transmitters) according
to the maximum output power of the communication equipment.
Separation distance according to the tranmission frequency (m)
Rated power of the
transmitter (W)
150 kHz to 80 MHz
d=1,2
√√√√
P
80 MHz to 800 MHz
d=1,2
√√√√
P
800 MHz to 2,5 GHz
d=2,3
√√√√
P
0,01 0,12 0,12
0,23
0,1 0,38 0,38
0,73
1 1,2 1,2
2,3
10 3,8 3,8
7,3
100 12 12
23