DUODYNATOR
®
1049
9
gbo
Medizintechnik
AG 2008
Version 1.3
1.5 Short instructions: treatment with interference currents
1.
Connection through the mains switch at the back of the device.
2.
The device executes an automatic check of all functions.
3.
The faultless automatic check ends with the acoustic signal.
4.
Connect the accessories for interferential currents with the according socket at the side of the device
you wish to treat the patient. Note: Both sides have identical functionality and can be used
simultaneously! If necessary, switch on the connection cable of the
NEOSERV 1044
.
5.
Apply the electrodes on the patient.
6. Press
the
ICT key
on the side you wish to start the treatment.
7.
Open the interference programs menu by pressing the
Softkey
[Menu] on the same side.
8.
On the display the list of programs appears.
9.
Move the scroll bar/cursor with the
Modification knob
to the required interference program.
10.
Confirm by pressing the
OK key
.
11.
The required interference program is selected.
12.
At the side you wish to start the therapy, turn the
Intensity regulator
slowly, until the current is
clearly perceptible, but do not generate any muscular contracture, the therapy starts.
13.
If required, regulate the stimulation shifting. Activate the shifting regulator
by pressing the
respective
Softkey
. You may now regulate the stimulation shifting with the
Modification knob
.
14.
At the end of the therapy the triad gong will sound. The current to the patient will be automatically
reduced.
1.6 Short instructions: treatment with MFmod programs
(single circuit currents)
1.
Connection through the mains switch at the back of the device.
2.
The device executes an automatic check of all functions.
3.
The faultless automatic check ends with the acoustic signal.
4.
Connect the accessories for stimulation currents with the according socket at the side of the device
you wish to treat the patient. Note: Both sides have identical functionality and can be used
simultaneously! If necessary, switch on the connection cable of the
NEOSERV 1044
.
5.
Apply the electrodes on the patient.
6. Press
the
MF
mod
key
on the side you wish to start the treatment.
7. Open
the
MF
mod
programs menu by pressing the
Softkey
[Menu] at the desired channel.
8.
On the display the list of programs appears.
9.
Move the scroll bar/cursor with the
Modification knob
to the required program.
10.
Confirm by pressing the
OK key
.
11. The
required
MF
mod
program is selected. The channel is activated and shown inversely.
12.
At the side you wish to start the therapy, increase the current slowly with the
Intensity regulator
until
the desired stimulation effect occurs.
!!
Note!
The intensity can only be increased if there is an active channel. The selection is automatically
removed after 15 seconds if there is no user interaction. If you try to increase the intensity without
an active channel, the following note will appear:
Select
channel!