SELF-TEST, RED DOT SIGHT AND BLUETOOTH
Initialization /Self-test
The GAZOSCAN™ automatically performs a Self-Check when powered on. The initialization
Self-Check is performing a series of internal diagnostics to verify the integrity and performance of
the sensor and components. The device will continue to the subsequent user screens upon a
successful result of the diagnostics analysis. If any errors are found during the Self-Check, an error
screen will occur. In this case, contact your local GAZOMAT™ service center for assistance.
Instructions on the red dot sight
Users can look through the red dot sight to achieve better aiming. The red dot sight is powered by
the device. It automatically lights up after powering on the device.
NOTE: The sight is effective when aiming over 15m away. If the distance from the measured object
is too short, the red dot will not overlap with the green indication laser.
Red dot sight diagram
If the red dot and the green indicator laser do not overlap when aiming for a long distance (>15m),
the red dot calibration can be performed by adjusting the knob on the sight.
If the red dot is higher than the green
indication laser, turn the knob
counterclockwise to make it lower.
If the red dot is lower than the green indication
laser, turn the knob clockwise to move it up
.
If the red dot is to the right of the green indication laser,
turn the knob counterclockwise to shift it to the left;
If the red dote is to the left of the green indication laser,
turn the knob clockwise to shift it to the right.