CODE BLUE® MULTIPURPOSE |
User Guide
Options | 57
7.2 BLOOD PRESSURE LEFT ARM WITH OMNI®2 TABLET
FOR THE S303.250 ONLY
BP Cuff Calibration
1.
Follow the instructions in Section 7.1 "BP Arm
Setup" and connect the OMNI® 2 tablet.
2.
Tap the Menu icon, then select “Calibration”
3.
The Calibration screen will appear. Select
“Left Blood Pressure” to begin.
4.
Tap on START to begin Calibrating. Set
the BP cuff pressure to the values on
the screen starting with 0 mmHg.