46 | Working with OMNI Code Blue
CMR RATE
•
This submenu is for adjusting the number of
compression and ventilations per minute.
•
The default value for the compression
rate is 100 compressions per minute.
Adjust rate using the + and - buttons.
•
Press OK to save changes and
return to previous screen.
•
The compression rate limit is 50 - 150
CPM. the ventilation rate is 2 - 60 VPM.
CALIB.
•
This submenu is to calibrate the
pressure sensor in the simulator.
•
Press CALIB. to access additional
calibration menu selections.
•
See section 4.2 for instructions to
calibrate ventilations and section
4.3 to calibrate compressions.
RESET PRESSURE SENSOR
•
It is recommended that the OMNI® controller’s
pressure sensor is reset at altitudes
greater than 1000 ft. to avoid inaccurate
compression and ventilation readings.
•
Perform the sensor reset procedure as
part of the initial calibration process.
1.
Select RESET within the compression or
ventilation calibration menu.
After resetting the sensor, OMNI®
will display DONE.
2.
Select SAVE to finish