Camera Connector (Figure A5): Connections to the UltraScan camera
are via a 62 pin connector located on the rear of the controller.
Figure A 6
Controller’s back panel: air connections
•
Air Connections (Figure A6): The microscope air line at 60 PSI is
connected to the AIR “IN” connection located on the rear panel. The
camera insertion air line is connected to the AIR IN port. If an “inserted
camera” state is requested by the user, an electronic air valve connects
AIR IN to AIR OUT. If a “retract camera” state is requested or if the
controller is powered down, the AIR OUT port is connected to the
AUX port. No connections to the AUX port are necessary.