6. Detector Operation Flow
6.3. Operation Flow
■ After power on, it passes self-diagnostic process then enters Measuring Mode. Here, by operating front keys,
you can go to internal System Mode.
■ Timeout for Level1 and Level2 are 10 sec. It is set to 1 hour for Level2 Calibration and Test Mode.
■ When“RESET”key contacts at Program Mode Screen, it returns to Measuring Mode. When "RESET"
key contacts at each Program Setting Screen, it returns to the parent step.
[Figure 13. GTD-3000Ex Mode Configuration]
[Table 9. Operation Key Description]
ITEM
NAME
DESCRIPTION
FUNC
Function key
Detector Mode Setting Entry Function (Insert Magnetic-bar for 2 sec or longer in measuring
mode) Function to enter the next stage of Level2 and to save the setting
RESET
Reset key
Move to the previous stage from the level entered.
↑
Up key
Next level mode that is configured in LEVEL1 and Change in Level2 setting Plus
↓
Down key
Next level mode that is configured in LEVEL1 and Change in Level2 setting Minus
■ Operation keys for detector system mode are defined as below.