GTD-2000ExW
Instruction Manual
6.4. Menu Configuration Table
LEVEL1
LEVEL2
DEFAULT
NAME
PARAMETER
PROGRAM
MODE
GAS TYPE (Gas Type)
[DEFIN./USER]
DEFIN.
GAS SEL(Gas Select)
Built-in gas name selection
COMB.
UNIT SEL (Unit Select)
%/%LEL/PPM/PPB
%LEL
D-POINT(Decimal Point)
0.100/1.00/10.0/100
100
HIGH SCL (High Scale)
1~10000
100
PASSWORD
00~99
00
CALIBRA.
MODE
(Calibration Mode)
CALIBRA.
[ZERO]
(Calibration Zero)
ZERO CAL
[NO,YES]
[NO]
ZERO GAS
[ 0]
ZERO >>>>
[SUCCESS / FAIL]
CAL. DATA
[ 0/ FAIL]
CALIBRA.
[SPAN]
(Calibration Span)
SPAN CAL
[NO , YES]
[NO]
SPAN GAS
[ 0]
SPAN SET
[ 50/ FAIL]
50
SPAN >>>>
[SUCCESS / FAIL ]
CAL. DATA
[ 0]
ALARM
MODE
ALM USED (Alarm Used)
[OFF / ON]
OFF
AL LATCH (Alarm Latch)
[OFF / ON]
OFF
A1 LEVEL (Alarm-1 Level)
[1 ~ Full SCL]
20
A1 TYPE (Alarm-1 Type)
[INC / DEC]
INC
A1 DBAND (Alarm-1 Dead Band)
[0.0 ~ 10.0%F.S]
0
A1 TIME (Alarm-1 Time)
[0~ 60]SEC
1
A2 LEVEL (Alarm-2 Level)
[1 ~ Full SCL]
40
A2 TYPE (Alarm-2 Type)
[INC / DEC]
INC
A2 DBAND (Alarm-2 Dead Band)
[0.0 ~ 10.0%F.S]
0
A2 TIME (Alarm-2 Time)
[0~ 60]SEC
1
SENSOR MODE
Factory Setting
MAINTEN.MODE
Factory Setting
DEVICE MODE
Factory Setting
VERSION MODE
Factory Setting
TEST MODE
Factory Setting
7.1. Program Mode
PASSWORD
[**]
- Contacting "FUNC" key with the Magnet-bar for 2 sec or longer in Measuring Mode enters Password mode.
- After setting Password using "↑" or "↓" key, contact "FUNC" key.
PROGRAM
MODE
- If password is correct, it enters Program mode.
- By contacting "↑" or "↓" key, mode changes in defined order.
-
( PROGRAM -> CALIABRA. -> ALARM -> SENSOR -> MAINTEN. -> DEVICE -> VERSION -> TEST )
GAS TYPE
[ DEFIN. ]
- Mode that sets gas name type and contacting "↑" or "↓"key changes gas type.
-
(DEFIN. / USER)
- DEFIN. Is acronym for Define and is selected when using built-in setting of gas name. USER is
-
selected when the user directly sets a gas name.
- When a desired gas type is displayed, contact "FUNC" key to set the gas name and enter Program Mode.
GAS SEL
[ COMB. ]
- Mode that sets gas name and contacting "↑" or "↓"key changes gas name.
- When gas type is set to DEFIN., built-in gas names can be selected for use. When it is set to USER,
-
the user shall set gas name using 5 character for use. Characters allowed for use are number, capital
-
Alphabet, space, and period. When position is moved by setting characters, a cursor in black box shape
-
appears as shown in the figure on left.
USER GAS
[■SER ]
UNIT SEL
[ %LEL ]
- Mode that sets gas measuring unit and contacting "↑" or "↓"key changes gas measuring unit.
-
( % / %LEL / PPM / PPB ).
- When a desired gas measuring unit is displayed, contact "FUNC" key to set the unit and enter Program
-
Mode.
D-POINT
[ 100]
- Mode that changes decimal place and contacting "↑" or "↓"key changes decimal place.
-
(0.100/1.00/10.0/100)
- When a desired decimal place is displayed, contact "FUNC" key to set the decimal place and enter the
-
next Program Mode.
HIGH SCL
[ 100]
- Mode that sets high scale value that should be displayed for full range. Contacting "↑" or "↓"key
-
increases or decreases the scale value.
-
( 1 ~ 10000 )
- When a desired high scale is displayed, contact "FUNC" key to set the high scale value and enter the
-
next Program Mode.
PASSWORD
[00]
- It is password setting mode. This password is checked when entering Program Mode.
- Set using "↑" or "↓" key then contact "FUNC" key to set the password and enter the next Program Mode.
7. System Mode
6. Detector Operation Flow