- It is password change menu.

- The password is for entering device setting menu and pressing UP (▲) or DOWN (▼) key 

increases or decreases number for password.

- Password can be set in a range between '00~99 and default is '00'.
- After setting the password, press FUNC key to save the setting and move to the next menu.

- It is a function to automatically change measuring channel. When it is set to On, it changes 

from channel 1-1 to 16-8 in an order each sec and displays the measurement in measuring 

mode.

- Default setting is 'OFF'. Press FUNC key to save the setting and move to the next menu.

- It displays end of configuration menu. Pressing FUNC key changes to the initial menu page 

"[F1]/CONF".

www.gastron.com

18

_

19

GTC-700

Instruction Manual

■ When "FUNC" key is pressed for 2 sec or longer in measuring mode, it enters password input step.

6.3.1. Configuration Mode

- After entering password using UP (▲) or DOWN (▼) key followed by pressing FUNC (Function) 

key, it enters Function Mode. (Default = "00")

- It displays "FAIL" message in an event of password error.

- Using UP (▲) or DOWN (▼) key, select “[F1]/CONF” (Configuration menu) then press FUNC 

key to enter sub-menu setting.

- It is a menu that sets Modbus Slave Address.

- For address input, pressing UP (▲) or DOWN (▼) key increases or decreases number.

Setting range is OFF~32 and default is OFF.

- After setting the address, press FUNC key to save the setting and move to the next menu.

6. Operation Method

6. Operation Method

Summary of Contents for GTC-700

Page 1: ...ce d Industry 1 ro Bugok dong Gunpo si Gyeonggi do Tel 82 31 490 0800 Fax 82 31 490 0801 Yeongnam business office Plant 55 Gonghangap gil 85beon gil Gangseogu Busan Metropolitan City Tel 51 973 8518 F...

Page 2: ...ormed quarterly when the calibration period is not specified The product and manual are subject to change without a notice for the product s functional improvement and ease of use For accurate operati...

Page 3: ...onmental Specifications 7 4 Name and Description of Each part 8 4 1 External Components 8 4 2 Internal Components 10 5 Installation 11 5 1 Power and External Signal Configuration 11 5 2 Channel Config...

Page 4: ...e Display Common Unit 4 Digit Seven Segment x 2 Analog Unit Power Error Status LED Output Signal Isolation RS 485 MODBUS MODBUS TCP IP Cable Connection Length and Wire Requirement Power Approx 40 m wh...

Page 5: ...ights on when power turns on 16 Main Power Switch Power ON OFF Switch 17 Analog Input Channel Status LED It displays measurement status Orange light turns on below 3 9 mA Green light turns on between...

Page 6: ...24 V Connected to the external power supply V terminal for device power supply provides normal 24 V Connected to the external power supply FG Frame Ground terminal that is connected to the power suppl...

Page 7: ...125 125 CH 16 5 126 126 CH 16 6 127 127 CH 16 7 128 128 CH 16 8 Channel configuration per slot is as shown in the table below www gastron com 12_13 5 Installation 5 Installation GTC 700 Instruction Ma...

Page 8: ...e with CVVSB 0 7 sq or higher Check wirings for operation power detection parts alarm parts and concentration display Check the power input then turn ON the power S W Check whether power LED at alarm...

Page 9: ...w 4 mA green lights on for values in a range of 4 mA 20 mA and red lights on for values exceeding 20 mA Table 6 Operation Setting Mode Default Settings LEVEL1 LEVEL2 LEVEL3 DEFAULT F1 CONF Configurati...

Page 10: ...tion menu Pressing FUNC key changes to the initial menu page F1 CONF www gastron com 18_19 GTC 700 Instruction Manual When FUNC key is pressed for 2 sec or longer in measuring mode it enters password...

Page 11: ...an move between channels 1 1 16 8 for calibration Pressing RESET key moves to the previous step Using a current generator input 20 mA output into the analog input terminal positive negative of the cha...

Page 12: ...clockwise direction in an order For LED of analog input unit power status LED lights on and Error Com Channel Red Green Status LEDs flash in an order Pressing FUNC key ends the test and moves to the...

Page 13: ...nge during menu setting It displays the end of the selected menu After approx 1 sec it moves to the initial menu for network setting Pressing FUNC key for 2 sec or longer followed by password input en...

Page 14: ...r DOWN key to select between 0 255 Pressing FUNC key sets the selected address and moves to the next address setting menu It is the second subnet mask setting menu Use UP or DOWN key to select between...

Page 15: ...d moves to the next address setting menu It is the fourth gateway setting menu Use UP or DOWN key to select between 0 255 Pressing FUNC key sets the selected address and moves to the next address sett...

Page 16: ...lue between 00 FF hexadecimal When there is a change it returns to the selected menu To complete the change of setting use UP or DOWN key in the selected menu to select UP y If the change is not wante...

Page 17: ...ion Unsigned Integer Refer to section 9 Error code Heart bit 30191 Repeats 0 and 1 Unsigned Integer 30192 Increase by 1 Unsigned Integer Table 8 Fault List MESSAGE VALUE DESCRIPTION CONDITION RECOVERY...

Page 18: ...www gastron com 34_35 GTC 700 Instruction Manual Figure 8 GTC 700 Drawing VERSION CONTENTS DATE 1 0 Initial Revision of Manual 2016 10 05 10 Revision History 9 Drawings and Dimensions...