GasCheck 3000 & 3000is Ion Science Ltd
Page
9 of 23
2.3
Battery data
Note: The energy storage capacity of batteries is affected by temperature. The data in Table 2 is for an
ambient temperature of 20 °C (68 °F).
Number of batteries required
4
Battery type
Alkaline, size AA
Recommended batteries
LR6 (R6) or MN1500
Average operating life
see Table 2
Battery type
Battery life (hours)
Gas Check 3000
Gas Check 3000is
LR6 (R6) *
40
20
MN1500 *
40
20
R6S
20
10
R6PP
20
10
*
Preferred battery types
Table 2 – Average battery life
The use of the back light will dramatically shorten the battery life.
Do not change batteries in a hazardous area – see section 5.2.
WARNING
For GasCheck 3000is
Do not
modify the GasCheck 3000is equipment to allow any recharging.
For Intrinsic Safety T4 Rating in the Gas Check 3000is
rechargeable batteries (e.g. Ni-Cad) must not be used.
There is no provision for the recharging of batteries within the GasCheck 3000is.
For alkaline types (MN1500 or LR6) type use only Duracell (Standard, Procell or
Ultra)
or Ever Ready (Energiser) batteries. Alternatively use any zinc carbon types R6PP or
R6S
For GasCheck 3000
Do not
modify the GasCheck 3000 equipment to allow any recharging
Do not
use in a hazardous area. Use the GasCheck 3000is.
The GasCheck 3000 is designed for use with primary (disposable) batteries.
Rechargeable batteries can be used for the Gas Check3000 in safe areas.