17
raise or lower the readings to match
the concentration printed on the
calibration cylinder label.
DILUTION SPAN CALIBRATION
LEL
0
CO
0.0
H
2
S 0.0
SO
2
10.0
EXIT
7. When the span calibration for a
particular sensor is completed, press
the MODE button to confirm the
calibration and then use the
navigation arrows to advance to the
next sensor channel to be adjusted.
If using a calibration
balloon and the calibration of more
than one sensor is required, deflate
the calibration balloon fully before
filling it with new gas for subsequent
calibrations. Failure to fully deflate
the calibration balloon between
calibrations may lead to inaccurate
and potentially dangerous readings.
DILUTION SPAN CALIBRATION
LEL
0
CO
0
H
2
S
0
SO
2
0.0
EXIT
8. Repeat steps 5, 6 and 7 until all
sensors have been calibrated. When
finished, use the navigation arrows to
highlight EXIT and press the MODE
button to exit and save the new
calibration settings.
SAVE CHANGES
YES
NO Cancel
9. Press the MODE button with yes
highlighted to save changes.
Remove all fittings before turning the
instrument back on and returning to
normal operation.