12
3.3
Manual dilution calibration
procedures
Cannonball 3 instruments with dilution
capability have five different manual
calibration options in the Calibration
Menu: FRESH AIR, O2 ZERO, SPAN
CAL, DIL FRESH and DIL SPAN.
3.3.1 Fresh air calibration in dilution
mode
While in dilution mode, manual fresh air
calibrations include calibrations with and
without the dilution pump.
To perform a fresh air calibration:
1. Enter the Calibration Menu.
CALIBRATION MENU
GAS VALUES
SPAN CAL
HISTORY O
2
ZERO CAL
DIL. SPAN
FRESH AIR
DIL. FRESH
REMINDER
EXIT
2. Use the navigation arrows to highlight
FRESH AIR as shown above. Press
the MODE button to confirm the
selection.
Do Fresh Air Cal Now?
YES
NO
3. Press the MODE button with YES
highlighted to begin the fresh air
calibration.
The dilution pump will be automatically
disabled by the instrument during the first
stage of the fresh air calibration.
Fresh Air Calibration
Please Wait…. 14
After the completion of the fresh air
calibration without the dilution pump, the
instrument will re-enable the dilution
pump and automatically proceed to the
fresh air calibration with the dilution
pump.
Fresh Air Calibration
Please Wait…. 29
DILUTION PUMP ON
↓
Fresh Air Calibration
Completed
3.3.2 O2 zero calibration in dilution
mode
In dilution mode, the O
2
zero calibration
includes two separate calibration
routines. The first calibration routine
calibrates the oxygen sensor to zero with
the dilution pump disabled. The second
routine enables the dilution pump and
sets the O
2
zero point for use with the
dilution pump enabled.
The O
2
zero calibration adjustment in the
Cannonball 3 should not be confused
with the fresh air calibration adjustment
of the oxygen sensor. The O
2
zero
calibration adjustment is used to calibrate
the oxygen sensor in an atmosphere
containing no oxygen, while the fresh air
calibration adjustment calibrates the
oxygen sensor in an atmosphere
containing 20.9% Oxygen.
To perform the O
2
zero calibration while
in dilution mode:
1. Turn the Cannonball 3 on and wait at
least three minutes to allow sensor
readings to stabilize fully before
initiating any calibration procedures.
2. Enter the Calibration Menu.