1. Zero Point Calibration

Open the detector in clean air, if there is readings after counting down.

Press the button "MENU" of the remote controller, the detector displays "F-01", press"+" or"-", change 

the function to "F-04", and click "OK". The detector value will be displayed as 0.

In the zero calibration of oxygen detection transmitter, we must use nitrogen of purity over 

99.99%.

PRV

PTFE PIPE

Empty

Gas detector

Standard gas

Flowmeter

2. Range Calibration

1. Enter into "F-07"as the same way above . The

"F-07" value is the concentration of the standard gas, 

initialized to half of the measure range. This value can 

be modified through the"+" and"-" keys to make it 

consistent with the standard gas concentration which 

need to be calibrated. Press 

"OK" to exit.

2. Enter into"F-08"to change 0 into 1,then back to

detection interface.

3. Connect the 200-400mL/min standard gas in

detection interface, when the detector value is stable, 

press "OK" for 3 seconds, then the instrument will be 

calibrated automatically. The measured concentration 

value is consistent with the value set in "F-07".

4. After calibration, left the detector in the air for a

while. Shut off the power when the detector value is 

close to zero.

Common Malfunction, Repair and Maintenance

Product component

Fixed Gas Detector 

Instruction Manual 

Qualification Certificate 

Remote Control

one

one

one

one

Symptom

No output
for
transmitter

Reason
wrong connection

potential in reference point is not 

correct

Too low
host
reading

Circuit fault

Value of S potentiometer 

setting is too small

Too high
host
reading

Transmitter failure

Value of S potentiometer 

setting is too large

Transmitter failure

Instable
host
reading

Heat settling time is not 

enough

Transmitter failure 
Circuit fault

Method to process

connect again

Adjust zero point gradually and then adjust potentiometer (4mA) Return to 

Yuante for repair

Calibrate again

Get a new transmitter

Calibrate again

Get a new transmitter

Wait for starting up

Get a new transmitter

Return to Yuante for repair

www.gasdog.com

[email protected]

+86 773-280-9002

11

Summary of Contents for 41547

Page 1: ...brand sensors and advanced micro process technology with data display fast response high preci sion and good stability Powered by 24V DC with 4 20mA RS485 relay signal output Explosion proof suitable...

Page 2: ...he specific gas concentration in environment or pipeline and alarm With self calibration and zero calibration function makes the detection more accurate and reliable With temperature compensation this...

Page 3: ...0 25 VOL 0 2000ppm 0 1000ppm 0 100 VOL 0 200ppm 0 100ppm 0 100ppm 0 40000ppm 0 250ppm 0 100ppm 0 2000ppm 0 100ppm 0 1000ppm 0 50000ppm 0 5000ppm 0 1000ppm 0 5000ppm 0 100ppm 0 90 VOL 0 50ppm 1 LEL 0...

Page 4: ...Interior Decoration and Renovation Dyestuff Papermaking Pharmacy Health Care Foodstuff Antiseptic Disinfection Chemical Fertilizer Resin Adhesive Pesticide Raw Material Sample Technological Process L...

Page 5: ...iability if users operate the repair of replacement by themselves E There is acid solution inside the transmitter please do not disassembleit And pay attention not to damage the frontal membrane of th...

Page 6: ...tion Ground connection port 5 6 1 2 3 4 7 NC Normally close NO Normally open COM Common 24V 24 positive GND 24 negative 4 20mA 4 20mA analog signal Low Alarm High Alarm Relay Relay Powered Signal outp...

Page 7: ...2 F S H2 2 F S CO2 2 F S C2H4 2 F S NO2 2 F S N2 2 F S TVOC 2 F S NH3 2 F S ASH3 2 F S C3H3N 2 F S F2 2 F S O3 2 F S CL2 2 F S BF3 Range 0 30 VOL 0 1000ppm 0 1000ppm 0 20000ppm 0 10000ppm 0 100ppm 0 2...

Page 8: ...ress to modify the value Press OK to save settings and BACK to cancel 3 Continue to press to enter F03 maximum value setting menu of gas alarm Press OK to modify press to modify the value Press OK to...

Page 9: ...pment wind direction and annual meteorological conditions and even the location of windows and doors in an indoor environment We provide the following suggestions for your reference A Maintain a prope...

Page 10: ...iable ground wire For the wire between the transmitter and the main engine the shorter the better And protect the wire with iron pipe Method of Calibration To guarantee the measurement accuracy of the...

Page 11: ...seconds then the instrument will be calibrated automatically The measured concentration value is consistent with the value set in F 07 4 After calibration left the detector in the air for a while Shut...

Reviews: