Multi Gas Clip User’s Manual
UM MGC V2.0
8
12 of 21
•
When a bump interval has been programmed into the detector, but a bump test is not yet due
, the
current bump test status will show: “
L . BUMP
___” (the date of the last bump test) followed by “
BUMP ___d
.”
(the number of days until the next scheduled bump test is due) . Quickly press the power/menu button
while the display shows “
BUMP ___d
.” to force a manual bump test . Audible, visual & vibrating prompting
to apply gas will start . The display will alternate between “
BUMP DUE
” and “
APPLY GAS
” . Snap the MGC
Calibration Cap (provided with the detector) into place over the sensor ports . Apply gas to the detector at
a flow rate of 0 .25 to 0 .5 LPM and do not disturb while test is being performed . Once all of the sensors have
been tested, the detector will go into alarm . Remove the MGC Calibration Cap and the detector will return
to normal operation after a short period of time . The bump due date will be automatically reset based on
the new bump date . The bump test will be recorded in the Bump Log .
•
When a bump test interval has not been programmed into the detector
, the current bump test
status will show: “
L . BUMP
___” (the date of the last bump test) . Quickly press the power/menu button
while the display shows “
L . BUMP
___” to force a manual bump test . Audible, visual & vibrating prompting
to apply gas will start . The display will alternate between “
BUMP XXXd
.” and “
APPLY GAS
” . Snap the MGC
Calibration Cap (provided with the detector) into place over the sensor ports . Apply gas to the detector at
a flow rate of 0 .25 to 0 .5 LPM and do not disturb while test is being performed . Once all of the sensors have
been tested, the detector will go into alarm . Remove the MGC Calibration Cap and the detector will return
to normal operation after a short period of time . The bump test will be recorded in the Bump Log .
L. BUMP
L. BUMP
BUMP
d.
BUMP
XXXd.
Press power/menu
button
Press power/menu
button