Page 25
810-PS2 Manual
3. Select a memory location with MEM LOC.
4. Plug the programmer plug into the gauge display to be copied
from. Press the COPY button to copy the gauge calibration
into memory.
5. When the operation is complete, unplug the programmer plug
from the first gauge and plug it into the gauge display to be
copied to. Press the PROG button to transfer the calibration
to the second gauge.
6. When the operation is complete, unplug the programmer
from the gauge and verify gauge operation.
Program an 810-PS2 gauge display from a table of calibration
values:
1. Obtain a table of inches versus volume
2. Turn on the programmer.
3. Make sure the inch mode is correct.
4. Select a memory location with MEM LOC.
5. Press the CLEAR MEM button to erase any previous calibration.
6. Starting at the 0 inch value of volume calibration, use the 0
to 9, decimal, E, F, L, U, or BL buttons on the keypad to enter
the calibration. Press the ENTER button to store the value in
memory. When ENTER is pressed, the inches will go to the next
value. If you make a mistake, use the BACK button to erase
the entry, or if ENTER has already been pressed, use INCHES
to go back to that inch value and re-enter the correct value.
If more than 4 numbers are entered the previous ones will
scroll off the left of the display. If ENTER is pressed before any
numbers, nothing will happen.
7. After the table has been entered, use the INCHES buttons to
review the table to make sure it is correct. If a calibration
value is incorrect, simply re-enter it and press ENTER.
8. Continue with the same procedure as in Program the gauge
display into inches from step 5. When setting the bottom
reading, use the calibration table to look up the volume
corresponding to the distance to the middle of the float.
Use the
OFFSET
buttons to obtain this reading on the
CALIBRATION display.