Page 20
810-PS2 Manual
To set the density:
1. The display must be showing a valid reading in order to set
the density. If “no L” or some other error message is showing,
repair or connect the gauge before proceeding.
2. Press and hold the button on the side of the enclosure. After
about 7 seconds the display will show the current density
setting, for example, “C .90” indicates a current density setting
of 0.90 which is the default. Release the button at this point.
3. If no further action is taken, the display will revert to normal
operation after about 5 seconds with no change in the density
setting. This is useful if you just want to check the current
density setting.
4. To change the density setting, press and release the button
repeatedly until the correct density is shown. This must be
started before the 5 second time expires, otherwise start
again at step 2. The “C” for “current density” on the display
will change to “P” for “Program density” and the density will
increase from the current setting in 0.05 increments for each
button press. For example, if the current density is 0.90, then
the display will show “P .95”, “P1.00”, “P1.05”, “P1.10”, “P1.15”,
“P1.20”, then it will start over at “P .60”, “P .65”, and so on.
5. When the correct density is shown, stop pressing the button.
After 5 seconds the display will show “Stor” for 2 seconds,
indicating that the new density value has been stored.
6. The display will then return to normal operation.
To program which float is being used:
1.
Since the plastic and stainless steel floats have different
buoyancies, the display must be programmed with the type
of float used so the density correction will be accurate. This
only needs to be done once during installation.
2. The display must be showing a valid reading in order to
program the float type. If “no L” or some other error message
is showing, repair or connect the gauge before proceeding.
3. Press and hold the button on the side of the enclosure. After
about 7 seconds the display will show the current density
setting, for example, “C .90” indicates a current density setting
of 0.90 which is the default. Continue to hold down the
button.
4. Put a magnet next to the display face by the “G” in Garnet.