4. Illustration of keys

0-9

Keys for entering numbers

.

In the number-entering state, it is a key for decimal. When not in the number-entering state, it is
the printing key.

CE

In the number-entering state, it clears the entered numbers.
When it is not in the number-entering state, press and hold the “CE” key followed by
pressing the “UNIT W.T.” key and you are in the calibration mode.

TARE

In the number-entering state, press the “TARE” key, packaging pre-removal is provided for
the current entered value.
When it is not in the number-entering state, pressing the “TARE” key clears the current
weight.
In the calibration state, press the “TARE” key and you are out of the calibration state.

kg/lb

The unit in the weight window between kg(or g) and lb.

UNIT
W.T.

In the number-entering state, press the “UNIT W.T.” key and the current entry is set as unit
weight.

SMPL

In the number-entering state, press the “SMPI” key and the current entry is set as number of
pieces.
To set the number of pieces, the displayed data in the weight field must be more than 0.
Otherwise the set number of pieces is 0.

M+

When the displayed data in the weight field is more than 0, accumulated weight value and
value for number of pieces are re-established after the unit weight window displays the
accumulated number of operations for 5s, and the next accumulation is possible only when
the weight returns to 0.

MC

Allowing you to get out of the accumulation state and clear the accumulated value.

QTY
PST

In the number-entering state, press the “QTY PST” key to set the current entry as the upper
limit of the quantity.
When it is not in the number-entering state, the quantity window displays the preset quantity
and returns after 5s.

5. Information displayed on the counting weighing apparatus

5.1The weight field displays

F----1

Data exceeds the range of display.

F----2

Error in A/D conversion circuit

F----3

Indicating that keys are available

F----5

Indicating fault in the sensor

F----L

Zero of sensor signal is too low

F----H

Weight exceeds the range

C----F

Error in calibration. It is not at zero or the data are not stable during the calibration.

C----H

Error in calibration. Article is placed on the weighing platform or the zero of the sensor
is too high during the calibration.

E-----

Error in storing the calibration data. Service is possible only after re-calibration.

5.2 The unit weight field displays the following information:

-CAL-

You are in the calibration state

≡ X≡

The displayed value is the number of operations for accumulation. The weight field

Summary of Contents for TJ3K

Page 1: ...arm and more 1 5 Use of green LED display allowing distinct display and visible readings 1 6 The electronic counting weighing apparatus allows options of data output interface and printing button dire...

Page 2: ...he display window displays the negative value of the weight of the container Nevertheless press the TARE key and the Packaging Removal function is canceled and zero is displayed for the weight 3 3 Cal...

Page 3: ...perations and the quantity display window displaying the total weight After about 5s you are again in the counting state where the weight display window displays the current weight the unit weight win...

Page 4: ...ed weight value and value for number of pieces are re established after the unit weight window displays the accumulated number of operations for 5s and the next accumulation is possible only when the...

Page 5: ...Charging Lights when the counting weighing apparatus is in the charging state It takes about 8 hours for one charge Fully charged Slowly lights when the counting weighing apparatus is under charge Wh...

Page 6: ...display accumulation MW 12345 6 kg 15 bits MN 12 12 bits MQ 12345 Pcs 14 bits 7 3 Receive order need to receive the communication then receive order When the communication setting is 27 default statue...

Page 7: ...he producer offers a free repair service to all G G Product which become defect under proper usage without human damage The customers are kindly requested to mail the product as well as its original p...

Page 8: ...ical appliances are disposed of in landfills or dumps hazardous substances can leak into the groundwater and get into the food chain damaging your health and well being When replacing old appliances w...