Operating Manual for JJ-B Series of Electronic High Precision Balances

           (for following models: JJ100B, JJ200B, JJ1500B, JJ2000B)

1. Overview

        Manufactured and well-designed by G&G Measurement Plant, adopting 
industrial-class high precision sensors, measurement circuit, a “G&G” 
dedicated single-chip computer system, and controlling system, JJ-A series 
electronic balances are high-quality electronic weighing machines. Advanced 
technology, well-selected, exquisite processing and strict testing make JJ-A 
series of electronic balances be possession of following strong points:
        - Simple operations.
        - Integrated water level on the scale
        - High precision, hypersensitivity and fast reaction speed
        - High quality, strong anti-interference capability, long life, long-term 
using  stability,  adapting  to  the  adverse  environment  and  long  hours 
continuous work
        - With auto calibration
        - Unit conversion feature: in seven units between arbitrary conversions
        - Balance is targeting output data interface and can be connected directly 
to the printer and computer to collect and analyze the data. The balance can 
even remote controlled by computer. 
        In general, with high precision, high stability, high reliability and high 
efficiency, the “G&G” electronic balances has widely replace the traditional 
mechanical, tray balances. The “G&G” electronic balances have been widely 
used in all callings, in different kinds of measurement and test. Those have 
played an important role in improving products quality, cost-effectiveness 
and efficiency.

2. Specifications and Performance indexes

        Depending on the difference of range and resolution, G&G Brand JJ-B 
series electronic scales are classified to 4 types as shown 

 Model                                  JJ100B               JJ200B              JJ1500B             JJ2000B
 Capacity                                110g                   200g                 1500g                  2000g
 Resolution Ratio                  0.001g                0.001g                 0.01g                  0.01g
 Calibration Graduation          10d                      10d                    10d                     10d
 Tare Scope                           2000g                                            3000g 
 Calibrating Weight              2000g                                             2000g 
 Dimension of Platform                                             

Ф

135mm

 Volume                                                      325(W)X 211(D)X 111(H) mm
 Power Supply                          AC Adapter (AC 230V 15V 300 mA 50Hz-60Hz)
 Use Temperature                                                        0-40°C
 Use Humidity                                                         

80%R.H

        P.S. Only use the provided adapter or the same specifications of the 
adapter, otherwise there will be dangerous. 

3. Operational Method

        3.1 Turn on the power. Open the switch. The displaying windows 
displays from “F----0” to “F----9” and after a certain time for stabilizing,
 “0.0” appears; Then preheats for 15 minutes. Instability of the displaying 
at every beginning is normal; after a certain time, the displaying will stabilize.
        3.2 With empty scale platform, the displaying deviates from zero, 
pressing “TARE” key, then the displaying will return to zero.
        3.3 If the scale has not been used for a long time or just bought, the 
scale should be calibrated. Firstly, the scale is fully preheated (over 15 min.) 
with empty scale platform. Then press “CAL”, the displaying window shows 
“CXXX” and the balance will automatic calibrate (XXX should be the 
weight of standard weights). At the same time, only place calibrating weights 
on the scale platform. After stabilizing, the scale displays the weight of the 
calibrating weights; till now the calibration is completely finished and the 
scale can be normally used. If press “CAL”, but display “C----F”, which 
indicates the Zero point is not stable. Then re-press “TARE” to make “zero” 
point stable; and re-press the “CAL” to carry out calibration.

Summary of Contents for JJ100B

Page 1: ... JJ2000B Operation Manual G G GmbH Koenigstr 34 41460 Neuss Germany G G GmbH Content 1 Overview 2 Specification and technical data 3 Operation method 4 Attachment Configuration and description of function Change 5 Data transmission 6 Structure 7 Caution for Operation 8 Trouble shooting 9 Packing list 10 Disposal 11 Warranty 12 Contact ...

Page 2: ...es Depending on the difference of range and resolution G G Brand JJ B series electronic scales are classified to 4 types as shown Model JJ100B JJ200B JJ1500B JJ2000B Capacity 110g 200g 1500g 2000g Resolution Ratio 0 001g 0 001g 0 01g 0 01g Calibration Graduation 10d 10d 10d 10d Tare Scope 2000g 3000g Calibrating Weight 2000g 2000g Dimension of Platform Ф135mm Volume 325 W X 211 D X 111 H mm Power ...

Page 3: ...the con figuration and function the display shows C1 and the relevant number which can be changed Pressing CAL can change the parameters pressing TARE can change the values of parameters 4 1 C1 sensitivity setting 0 1 2 3 4 the larger number is the less sensible but the better stable Default setting is 2 or 1 4 2 C2 filtering strength setting 0 1 2 3 the larger number is the lower reaction speed b...

Page 4: ...ectronic balance without power supply 7 3 Place the scale on a firm table for operation and make sure that the operating environment of the balance is free of great vibration and interfe rence of power supply and of corrosive gases and liquids 7 4 Ensure the warm up time after power on 7 5 Warning If the equipment is used in a manner not specified by this instruction the perfection provided by the...

Page 5: ... 41460 Neuss Germany Tel 49 2131 609 7271 Fax 49 2131 609 7273 Email info gandg de Website www gandg de Remarks Option by JJ100B JJ200B Quantity 1 1 1 1 1 Article Description Operation manual Scale pan Power cable Calibration weight Wind shield restore the fabric configuration 1 remove the object 2 restore the fabric configuration restore the fabric configuration Press the TARE key with empty pan ...

Reviews: