background image

6

5

           1BH+71H: calibration (like “CAL”)      
           1BH+73H: Units change
           1BH+74H: Tare
        Sampling demo, connection method of balance and computer, please 
download from our website: http://www.gandg.de 

6. Structure

1. Glass Windshield                                  2. Scale pan
3. Part from Windshield                            4. ON/OFF key
5. TARE key                                             6. Multi-function key
7. CAL key                                                8. Water level
9. RS 232 interface                                   10. Power supply

7. Cautions for operation

        7.1 Using environment: indoor, max. Altitude: 2000m, temperature: 
0-40°C, humidity 

80%, Power fluctuations less than 

±

10%

        7.2 The electronic scale is a precision instrument which needs to be 
gently handled during the weighing of articles. Avoid impact and heavy load 
on top (never exceed its max. weighing limit). Any forms of impact or 
overloading many cause permanent damage to the electronic balance, even 
in the use of electronic balance without power supply. 
        7.3 Place the scale on a firm table for operation, and make sure that the 
operating environment of the balance is free of great vibration and interfe-
rence of power supply, and of corrosive gases and liquids. 
        7.4 Ensure the warm-up time after power-on..
        7.5 Warning: If the equipment is used in a manner not specified by 
this instruction the perfection provided by the equipment may be impaired
        7.6 Gentle handling in daily using, with dry and clean cotton cloth 
cleaning.

8. Trouble Shooting

Measures to take

1.the ON/OFF switch must be changed

2. please contact our service 

department

please contact our service department

please contact our service department

please contact our service department

Possible causes 

1. The ON/OFF switch is  

defect

2.Internal Circuit is defect 

1.A/D-switch problem

2.Sensor is defect

Keyboard defect

1.Sensor is defect

2.A/D switch problem

Problems 

Display shows 

nothing

The message “F-2”

The message “F-3”

The message “F-5”

Summary of Contents for JJ100B

Page 1: ... JJ2000B Operation Manual G G GmbH Koenigstr 34 41460 Neuss Germany G G GmbH Content 1 Overview 2 Specification and technical data 3 Operation method 4 Attachment Configuration and description of function Change 5 Data transmission 6 Structure 7 Caution for Operation 8 Trouble shooting 9 Packing list 10 Disposal 11 Warranty 12 Contact ...

Page 2: ...es Depending on the difference of range and resolution G G Brand JJ B series electronic scales are classified to 4 types as shown Model JJ100B JJ200B JJ1500B JJ2000B Capacity 110g 200g 1500g 2000g Resolution Ratio 0 001g 0 001g 0 01g 0 01g Calibration Graduation 10d 10d 10d 10d Tare Scope 2000g 3000g Calibrating Weight 2000g 2000g Dimension of Platform Ф135mm Volume 325 W X 211 D X 111 H mm Power ...

Page 3: ...the con figuration and function the display shows C1 and the relevant number which can be changed Pressing CAL can change the parameters pressing TARE can change the values of parameters 4 1 C1 sensitivity setting 0 1 2 3 4 the larger number is the less sensible but the better stable Default setting is 2 or 1 4 2 C2 filtering strength setting 0 1 2 3 the larger number is the lower reaction speed b...

Page 4: ...ectronic balance without power supply 7 3 Place the scale on a firm table for operation and make sure that the operating environment of the balance is free of great vibration and interfe rence of power supply and of corrosive gases and liquids 7 4 Ensure the warm up time after power on 7 5 Warning If the equipment is used in a manner not specified by this instruction the perfection provided by the...

Page 5: ... 41460 Neuss Germany Tel 49 2131 609 7271 Fax 49 2131 609 7273 Email info gandg de Website www gandg de Remarks Option by JJ100B JJ200B Quantity 1 1 1 1 1 Article Description Operation manual Scale pan Power cable Calibration weight Wind shield restore the fabric configuration 1 remove the object 2 restore the fabric configuration restore the fabric configuration Press the TARE key with empty pan ...

Reviews: