Appendix A: Default Controller Configuration
31
Changes that Gamry Instruments Has Made to Default Settings
Set Omega Protocol, Command Mode, No Line Feed, No Echo, Use <CR>
Set Input Configuration, RTD 3 Wire, 100 ohms, 385 Curve
Set Output 1 to PID Mode
Set Output 2 to On/Off Mode
Set Output 1 On/Off Configuration to Reverse, Dead Band 14
Set Output 2 On/Off Configuration to Direct, Dead Band 14
Set Display to FFF.F degrees C, Green Color
Set Point 1 = 35 degrees C
Level
2
Level
3
Level
4
Level
5
Level
6
Notes
SoAk
Guaranteed Soak: soak time always preserved
CYCL
Guaranteed Cy
cle: ramp can extend but cycle time can’t
Note: tIM.F does not appear for 6 digit display that use a HH:MM:SS format
tIM.F
MM:SS
“Minutes : Seconds” defau
lt time format for R/S programs
HH:M
M
“Hours : Minutes” default time format
for R/S programs
E.ACt
StOP
Stop running at the end of the program
HOLd
Continue to hold at the last soak setpoint at program end
LINk
Start the specified ramp & soak program at program end
N.SEG
1 to 8 Ramp/Soak segments (8 each, 16 total)
S.SEG
Select segment number to edit, entry replaces # below
MRt.#
Time for Ramp number, default = 10
MRE.#
oFF
Ramp events on for this segment
oN
Ramp events off for this segment
MSP.#
Setpoint value for Soak number
MSt.#
Time for Soak number, default = 10
MSE.#
oFF
Soak events off for this segment
oN
Soak events on for this segment