Chapter 6: Adding a new Load/Power Supply to Framework
6-14
Key Name
Type
Required
Description
CalCellValue
Float
YES
This is the calibration cell value, in ohms.
CalCell
String
YES
This is the description of the calibration cell which is
used to tell the user what connection to make.
VchRange
Float
YES
This is the voltage, in volts, used to set the Vch range on
the IFC1010 instrument. For the calibration, the range is
usually set to the x1 gain or the 3 V range.
IchRange
Float
YES
This is the voltage, in volts, used to set the Ich range on
the IFC1010 instrument. For the calibration, the range is
usually set to the x1 gain or the 3 V range.
IApplied1
Float
YES
This is the first current, in Amps, applied to the cell. In
the case of a bipolar power supply, this should be a
negative current. In the case of an electronic load, this
will be a positive current, selected to be less than
IApplied2
.
IApplied2
Float
YES
This is the second current, in Amps, applied to the cell.
In the case of a bipolar power supply, this should be a
positive current. In the case of an electronic load, this
will be a positive current, selected to be higher than
IApplied1
.
TApplied
Float
YES
This is the time, in seconds, that each IApplied is applied
to the cell.
TRest
Float
YES
This is the time, in seconds, that the calibration routine
waits between
IApplied1
and
IApplied2
. This is to
help alleviate any temperature buildup in the calibration
resistor.
Enabled
Integer
YES
The enabled field is used to tell the calibration routine if
this calibration is active for the power device. If the
calibration is disabled, the previous values are not used.
0 = Disabled
1 = Enabled
PreciseCalCellValue
Float
NO
This is the precise value for the calibration resistor to be
used. This value will be printed on the resistor. The user
will have a chance to enter this value during calibration.
It is stored in the file once entered to simplify future
calibrations.
Summary of Contents for LPI1010
Page 2: ...ii...
Page 6: ......
Page 14: ......
Page 30: ...Chapter 4 Hardware Installation 4 10...
Page 36: ......
Page 62: ......
Page 64: ......
Page 66: ......
Page 67: ...Appendix D Certificate of conformity Appendix D VII Certificate of conformity...
Page 68: ...Appendix D Certificate of conformity Appendix D VIII...