Chapter 4: Hardware Installation
4-5
Connect Current Monitor and Current Control Cables
The LPI D-sub module has to cables connections with one directly leading to the LPI Cable End Module. The second
cable has two male BNC connectors which mate with the appropriate connectors on the electronic load or
programmable power supply as shown in Figure 4-5.
Figure 4-5
Schematic overview of current control (I_CTRL) and current monitor (I_MON) cable installation
The red-marked BNC connector lead is the current control (I_CTRL) output of the LPI1010 and goes to the mating
current control input of the electronic load or programmable power supply. The black-marked BNC connector lead
serves as current monitor (I_MON) of the LPI1010 and connects to the current monitor output of the electronic
load or programmable power supply.
Connect Voltage Sense Cables
The LPI1010 system includes two additional LPI Voltage Sense Cables (P/N 985-00211 and 985-00212) with
protected male banana connectors on each cable end, see Figure 4-6.
If one or both connectors on the electronic load or programmable power supply is not a female
BNC, use an appropriate adapter such as those shown in Appendix B of this manual.