The instrument does not replace a cardiac examination, but
serves to detect pulse irregularities at an early stage.
4. Memory function
Semi-automatic Blood Pressure Monitor
Gamma
TM
Model M-S automatically stores
the result measured.
To see the result of the last measurement
press and hold the START/MEMORY button
for at least seconds.
If an irregular heartbeat interval was de-
tected in the measurement stored in memo-
ry, the icon will be displayed.
5. error Messages/Troubleshooting
If an error occurs during a measurement,
the measurement is discontinued and a cor-
responding error code is displayed (picture:
error no. )
error No. Possible cause(s) / Solutions
ERR
The systolic pressure was determined, but then
the pressure in the cuff went down to less than
0 mmHg (diastolic blood pressure can’t be meas-
ured). The tube may have loosened after the systo-
lic pressure has been determined. Pulse could not
be detected.
ERR
Unnatural pressure impulses influence the meas-
urement result. Reason: the arm was moved during
the Measurement (Artefact)
ERR 5
The measuring signals are inaccurate and no re-
sult can therefore be displayed. Read through the
checklist for performing reliable measurements
and then repeat the measurement.*
HI
The pressure in the cuff is too high (over 00
mmHg) OR the pulse is too high (over 00 beats
per minute). Relax for 5 minutes and repeat the
measurement.*
LO
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.*
*Please consult your doctor, if this or any other problem oc-
curs repeatedly.
Other possible errors and their solutions
If problems occur when using the device, the following points
should be checked:
Malfunction
Remedy
The display remains
blank when the device
is switched on although
the batteries are in
place.
. Check battery installation/ polar-
ity.
. Remove the batteries and if the dis-
play is unusual, then exchange them
for new ones.
The pressure does not
rise although you are
inflating the bulb.
Check the connection of the cuff tube
and connect properly.
Summary of Contents for M1-S
Page 12: ...22 23 1 M1 S 1 1 M1 S 4 1 2 M1 S 2 2 1 1 2 1 5 1 2 START START...
Page 13: ...24 25 2 2 3 5 3 1 1 2 3 2 3 4 5 2 3cm...
Page 14: ...26 27 15 10 6 2 3 2 1 Start 2 0 2 40 160 180 3 3 3...
Page 15: ...28 29 3 3 4 20 300 3 5 Err 3 6 IHD IHD IHD...
Page 16: ...30 31 ERR 1 20 ERR 2 4 GammaTM M1 S START 3 5 2 ERR 5 HI 300 200 5 LO 40 1 2...
Page 17: ...32 33 1 2 3 15 6...
Page 20: ...38 39 50 140 120 9 3 5 20 10 6 2 8 8 14 350 150 50 2 4...
Page 21: ...40 41 M1 S Gamma Gamma 40 30 4 9...
Page 23: ...44 45 2 2 1 1 2 1 5 1 2 START 2 2 3...
Page 24: ...46 47 5 3 1 1 2 3 2 3 4 5 15 10 2 3c 6 2 3 2 1 START 2 0 2 40 160 180 3...
Page 25: ...48 49 3 3 3 3 4 20 300 3 5 Err 3 6 HD HD...
Page 26: ...50 51 ERR 1 20 ERR 2 ERR 5 HI 300 200 5 LO 40 HD 4 Gamma M1 S START 3 5 2...
Page 27: ...52 53 1 2 1 2 3 6...
Page 30: ...58 59 50 140 120 9 3 5 20 10 6 2 8 8 14 350 150 50 2 4...
Page 31: ...60 40 30 4 9...