SpectralLED® RS-7 USER MANUAL
Page 70
4.7
Error Codes and Alarm Messages
4.7.1
Error Codes:
?01 - missing argument
?02 - argument out of range
?03 - unrecognized command
?04 - buffer overflow
Exceeded the RS-7 input buffer size of 8192 bytes when a command is being received
?05 - LSQ fault
FTS failed to find a solution
?06 - channel power unreachable
Attempt made to command one or more channels to >100% power
?07 - channel CPA*power unreachable
The requested RS-7 output would have been reachable, except the CPA coefficients pushed
one or more channels >100% power
?08 - channel CPA*FBG*power unreachable
The requested RS-7 output would have been reachable even with the CPA coefficients
applied, except the initial attempt to set the output power level resulted in the
optical feedback pushing one or more channels >100% power before a stable output
state within STP% of PDT could be reached
?09 - channel power ADP limit
The requested RS-7 output is attempting to drive one or more channels past the allowable
deviation from nominal of ADP%
?10 - channel power SLM soft limit
Attempt made to command one or more channels past the soft limit SLM%
?11 - excessive feedback correction indicated
While attempting to output the requested RS-7 output, an excessively out-of-range
photodiode response was encountered
?12 - data ended unexpectedly early
When transferring target spectrum (TSP) spectral data to the RS-7, the host did not
provide the expected number of spectral data points
?13 - tristimulus will not converge
CCS failed to find a solution
?14 - invalid units, must be radiometric (0) or photometric (1)
STS cannot use UNI = 2
?15 - TSP is zero
Attempt made to scale target spectrum TSP via STS command, when target spectrum is all