SpectralLED® RS-7 USER MANUAL
Page 41
4.2.10
– OCL - Offending Channels List
In response to an error message such as “channel power unreachable” or “SLM exceeded”, you can ask the RS-7
for a list of those channels which were attempted to operate beyond their limits, or otherwise generated an
error. The list shows each offending channel along with the requested power level (as
UNI
2
, 0 - 100%) which
could not be accommodated.
Syntax:
OCL
request Offending Channels List
Examples:
uni1
select photometric units (lux or nits, depending on
CTY & IRR
)
Ok
scp0,10
set all channels to 10 lux/nits
?06 - channel power unreachable
one or more channels could not be set to 10 lux/nits (such as IR LEDs)
ocl
request offending channels
17,154.4
channel #17 wanted to go to 154.4% power to satisfy the request
23,5911
channel #23 wanted to go to 5911% power
extra <CR><LF> terminates list