SpectralLED® RS-7 USER MANUAL
Page 53
Examples:
cty
request current calibration type
I
RS-7 is operating as an Irradiant source
uni0
set units
UNI
to irradiance/radiance
Ok
out
report current output level
272.12
µW/cm
2
or µW/cm
2
/sr
uni1
set units
UNI
to illuminance/luminance
Ok
out
519.852
lm/m
2
(lux) or cd/m
2
(nits)
out300
set output to 300 lux (lm/m
2
)
Ok
Note that changing the output level using
OUT
can result in a chromaticity shift in the final output spectrum
OSP
, as
OUT
merely adjusts all channels’ power level to match the requested setting without regard to any
wavelength shift caused by the change in drive current. To maintain the current chromaticity, use
OUTC
instead.
4.5.3
- OUTC - Adjust integrated output power level of OSP as UNI while maintaining chromaticity
Adjusts the RS-7’s integrated output power level using the units set by
UNI
, while compensating for the LED’s
wavelength shift at the different drive current by adjusting the spectrum to maintain the same chromaticity.
When setting the output level, the “Ok” response will not be returned until the RS-7 has settled to an accuracy
of
STP
percent.
Syntax:
OUTC
[level]
set the output level to
UNI
units while maintaining chromaticity
Examples:
oxy
report current output spectrum
OSP
chromaticity
0.4481,0.4077
chromaticity is (0.4481,0.4077)
uni1
select photometric units
Ok
out
report current output power in photometric units
7858.5
unit is outputting 7858.5 lm/m
2
or cd/m
2
out4000
change output level to 4000 without regard to chromaticity shift
Ok
oxy
get chromaticity at the new power level
0.4494,0.4078
it has changed because of LED wavelength shift at the new power level
outc4000
this time ask it to maintain chromaticity
Ok
oxy
get chromaticity again