Unlocking of axes:
Each setting housing of the CPV has four
screws. Two screws are for adjusting and shift-
ing the Vernier scale in horizontal (X) and verti-
cal (Z) direction. Additionally, there are two fix-
ing screws (A and B) to fix the axes inde-
pendently of one another.
The transversal axis (Y) can be opened with
clamp C.
Zero setting:
Adjust all scales of the CPV to zero.
As appropriate check the zero-position with the
Reference calibration key. If the calibration
does not show satisfying results (clearly visible
gap during examination with calibration key) it is
necessary to adjust the device exactly and fol-
lowing the instructions (see below).
In case the results of the adjustment of the unit
are not satisfying despite precise setting of the
scales, it is possible - with the calibration key
closed - to unscrew the lateral adjustment
plates with an Allen key (included in delivery)
and then newly adjust and fix them.