Reference SL Articulator and Face Bow Systems
- 25 -
Adjust the face bow depth measured with the
reference point indicator P on the reference
point indicator T (with the silver adjustment
ring). The reference point indicator T is attached
to the face bow by means of magnetic retention.
When working with the Condylograph
comfort
Set I, the individual reference point indicator is
used instead of the magnetic reference point in-
dicator T. Move the individual reference point in-
dicator into position T and calculate the depth to
set by dividing the depth in position P by factor
1.15. This value must be set on the reference
point indicator without changing the face bow
height that was measured on the patient. Sub-
sequently, fixate the reference point indicator it
on the face bow using the upper clamp screw.
Lower the upper section of the transfer stand by
turning the adjustment screw, until the refer-
ence plane adjustment reed slightly touches the
anterior ball of the reference point indicator.