Reference SL Articulator and Face Bow Systems
- 21 -
With the GAMMA Condylograph
comfort
, you
can localize both the right and left hinge axis
points directly on the patient and mark them on
the writing labels stuck to the red flags (mandib-
ula in Reference Position!).
In addition, the orbital reference point is marked
with the reference point indicator. When using
the Condylograph
comfort
Set, this is carried
out with the reference point indicator P that has
the blue adjustment ring.
In case of the Condylograph
comfort
Set I, the
individual reference point indicator in position P
is used instead.
With these three points (anterior reference
point, right and left hinge axis points) you can
create a reference plane with a zero-line (hinge
axis in Reference Position), which will be trans-
ferred into the articulator as follows: