7
Protection against electric shock (5.1):
The Game Ready System is considered to be Class I (protective earth) when connected to
the Ault Model MW 128 power supply, and Class II (double insulated) when connected to
the GlobTek Model GTM-21097-5012 power supply.
Protection against harmful ingress of water (5.3):
This product provides ordinary protection against ingress of water.
Degree of safety in the presence of
fl
ammable anesthetics or oxygen (5.5):
Not suitable for use in an oxygen enriched environment or in the presence of
fl
ammable
anesthetics.
Electromagnetic interference:
This equipment has been tested and found to comply with the limits for medical devices
in IEC 60601-1-2:2001. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However,
there is no guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to other devices, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different
from that to which the other device(s) are connected.
• Consult the manufacturer or
fi
eld service technician for help.
D E V I C E C L A S S I F I C A T I O N
Powered by Direct Current
Type BF Applied Parts
Attention: Consult accompanying documents before use.
Manufacturer
Summary of Contents for 550450
Page 4: ...4...