6 ADVANCED CONTROLS
6.1 Advanced Channel Controls
Advanced Channel Controls for each channel provide options to change the needle type for the
selected channel, link two channels together, and program multiple freeze-thaw cycles.
6.1.1 Select Needle Type Control
1. To change the needle type for a channel, press and hold the
Channel
button to enter Advanced
Channel Controls for that channel.
2. Select the proper needle type from the drop-down menu (Screen 6-1).
3. Press the
OK
button.
Screen 6-1. Select Needle Type
6.1.2 Link Channels Control
1. Press and hold the
Channel
button to enter
Advanced Channel Controls
for that channel.
2. Press the
Link
button to link two channels together for simultaneous operation. When two
channels are linked, the
Channel
button shows both channels (Screen 6-2).
NOTE:
This function is not available on the channel labeled
ALL
. You can only link channels that
are on the same horizontal plane on the Needle Interface (e.g., 1 and 2, 3 and 4, 5 and 6).
Summary of Contents for Visual-ICE MRI
Page 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Page 20: ...1 10 User Manual This page intentionally left blank...
Page 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Page 56: ...4 14 User Manual This page intentionally left blank...
Page 66: ...5 10 User Manual This page intentionally left blank...
Page 80: ...9 2 User Manual This page intentionally left blank...
Page 118: ...12 2 User Manual This page intentionally left blank...
Page 120: ...13 2 User Manual This page intentionally left blank...
Page 121: ......