Structure of instrument
Measuring
The area to be measured on the work piece must be properly cleaned
in
order to avoid that the tracer getting dirty. If the roughness is measured
manually, the tracer is laid onto the work piece and held steadily. The green
indicator LED on the instrument must be on, confirming that the tracer and
the instrument are in condition ready to measure. After pressing the start key,
the measuring procedure is effected automatically. Shortly after, the measu-
ring values can be read on the LCD-display and printed out.
Pressing any key can interrupt a current measuring operation and bring
the tracer to a halt. After a few seconds the trace rgo back to its starting
position again.
Keyboard
Tracer
Remove the supporting shoe by using the enclosed hexagon key or by means
of the knurled nut (type VHF).
Insert the tracer into the traverse unit so, that the point looks to the prism.
Warning: By no means is the tracer be inserted forcefully; it must be held
with the utmost care at its largest diameter!
If manual measurements are to be effected, remount the supporting shoe to
the traverse unit once again.
The position of the supporting shoe is correctly reached when the work piece
is supported on its whole length by the traversing unit with supporting shoe.
If the measuring area is too small, the bottom surfaces of traversing unit and
supporting shoe must be on the same level.
Connections
On one side of the instrument there are three connections, one for the power
supply, one for the cable to the motor unit, respectively to the tracer, and
another one for the serial connection to a computer.
Calibration
1. Press any button, after some seconds, the instrument is ready
for operation
2. Press button ‚M’ in the display appears the menu
3. Press the button ‚R’ four times to Calibration
4. Press the button once
5. Press the button ‚R’ once to Reference
For the calibration, the cutoff 0.8 mm is set automatically!
6. Press the button ‚Lt’ or ‚R’, if you do not calibrate with the original
metal calibration piece.
7. Press the button once
8. Press the button ‚R’ once to cal Ra/Lt
9. Press the button once and confirm again with
10. The calibration procedure (4 measurements) lasts one
minute approximately.
11. Press the button ‚M’, till you are again in the measure mode
Setting of language
1. Press any key – the instrument will be ready for operation within
a few seconds
2. Press button ‚M’ in the display appears the menu
3. Press the button ‚R’ twice to configuration
4. Press the button once
5. Press the button ‚R’ four times to system
6. Press the button once
7. Press the button ‚R’ once
8. Press the button several times, the further languages will appear
one after the other.
9. After having reached the requested language, press the button ‚Lt’ once
10 Press the button ‚M’, till you are again in the measure mode
1 Mains adapter
7 Interface USB
2 Paper roll cover
8 Connection cable
3 Registration paper
9 Reset instrument
4 Paper roll compartment
10 Motor unit VH (VHF)
5 Standard tracer
11 Remote control
6 Supporting shoe
12 Supplementary tracer
Mains
adapter
USB
Motor unit
Key
Measuring mode
Menu mode
Switch-on/off instrument
Confirmation of selected menu
setting
Menu for instrument configu-
ration
Switch-on/off menu
Selection of Cutoff
Cursor to the right
Selection of length of traverse
Menu selection next menu
option
Start of measuring operation
Interrogation of measuring
values
Menu selection previous menu
option
Printout measuring values,
profile, tp
Cursor to the left
FUTURO SURFACE
The complete manual is onto the CD, which is delivered with the instrument.
1
2
7
3
8
4
9
5
10 11
6
12