Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The HealthGuard-15 is intended for use in the electromagnetic environment specified below. The
customer or the user of the HealthGuard-15 should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment – Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood,
concrete or ceramic tile. If
the floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient / Burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Not Applicable
The HealthGuard-15 does
not use AC Power / Power
Supply Lines during
operation
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Not Applicable
The HealthGuard-15 does
not use AC Power / Power
Supply Lines during
operation.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0,5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for 5 s
Not Applicable
The HealthGuard-15 does
not use AC Power / Power
Supply Lines during
operation.
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
The HealthGuard-15 was
tested and performs
normally within the
specification limits.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.