Distance Unit
Choose the preferred unit to display the measuring
results.
0.000m
0.00m
0.0 in
In 1/32
0’ 00” 1/32
0.000
0.00
0.00 ft
Angle Unit
Choose the preferred unit to display angles.
° (degrees)
%
Calibration
NOTE: Calibration function may affect precision of the
device! This item cannot be adjusted in default mode.
User need to follow below steps to enter calibration.
×
Turn off the device
×
Hold the Records / Save button [11] and press the
measure / ON [3] button shortly. The device will
start-up.
×
Release the Records / Save button [11] after the
start-up screen has disappeared. The mainscreen
is visible.
×
Enter the menu and navigate to Calibration like
described above. Now the Calibration function is
free to enter.
- 0.009m ~ +0.009m
USAGE
§
Reference point
Four different reference points can be used.
Front
Middle (screw)
Back
End-piece
Different reference points can be chosen using the Reference point / Timer button [10]. The
Reference point setting [a] will be visible on screen.