8. This device is designed for indoor use only. Do not expose it to rain or humidity.

9.  Caution should be used when treating areas of the skin that lack normal sensation (e.g. diabetic 

patients). Note: watch out for skin irritations. Pause the treatment if necessary. 

10.  High currents do not lead to a more successful treatment, but can lead to skin irritation  

instead. The current strength should be adjusted to the comfort of the patient. In cases of skin 

irritation, stop the treatment and seek advice from your doctor.

11.  If the treatment feels uncomfortable, reduce the current strength to a comfortable level or 

stop the session.

12.  If the treatment is ineffective or unpleasant, discontinue the sessions and seek advice from 

your doctor.

13. Before operation, the patient must sit still and all cables must be plugged in firmly. 

14.  The mains plug is specified as a part of this medical equipment. Only use the adaptor supplied 

by the manufacturer.

15. Do not submerge the device in water. 

16. Dropping the device may cause damage.

Side  effects / accompanying  symptoms

•  Skin irritations are one possible side effect of the intended usage. The treatment should be  

stopped until the cause of the irritation can be determined.

•  You will perceive every treatment differently. For this reason, the milliampere value should only 

be used as a rough guidance. The current strength should be adjusted according to your personal 

comfort level.

•  Iontophoresis treatment stimulates the circulation, so a harmless reddening of the skin after the 

treatment may occur.

•  During the treatment you may experience tingling or stinging sensations. In this case, you should 

reduce the current strength to a comfortable value.

• As a consequence of the treatment, the skin will become increasingly drier. 

•  If the treatment leads to skin that is too dry, chapped or cracked, stop the treatment and consult 

your doctor. 

Summary of Contents for SweatStop DE20

Page 1: ...Tabwater Iontophoresis Device SweatStop DE20 EN Instruction Manual...

Page 2: ...cation 11 Basics 11 Preparation 12 Hand treatment 13 Foot treatment 15 Combined treatment of hand and feet 16 Underarm treatment 18 Treatment interval 19 Power supply 20 Monitored battery accumulator...

Page 3: ...nt is the intended operator of the iontophoresis device under the instruction or guidance of the clinician The DE20 is intended to be used in medical facilities and in private homes by healthcare prof...

Page 4: ...function of the equipment might deteriorate 8 Do not use the iontophoresis device near the active high frequency equipment and the RF shielded room of an MRI scanner where the intensity of disturbance...

Page 5: ...plugged in firmly 14 The mains plug is specified as a part of this medical equipment Only use the adaptor supplied by the manufacturer 15 Do not submerge the device in water 16 Dropping the device ma...

Page 6: ...arger Charger for series of 4 NiMH accumulators Order number A 0011 Cables for electrodes Order nomber Z 0066 Tubs for tap water Order number Z 3000 Electrodes in sponges for the treatment of hands an...

Page 7: ...ller 2 Diverter for type of current 3 Outlet sockets for electrodes 4 LED indicator weak battery 5 Digital display 6 LED indicator empty battery 7 Charging socket 8 Battery compartment 9 Code control...

Page 8: ...in the cables according to their colours to connect the device with the electrodes Change the polarity each time you start the therapy unless otherwise recommended by your doctor The pole reversal can...

Page 9: ...apacity of the batteries and the charger new cells standard equipment 4x 2500 mAh take approximately 11 12 hours to charge 9 Code control The number of all completed treatments will be saved This code...

Page 10: ...he real current value Disconnecting the device from the current supply 240V 110V When the DC plug of the charger is plugged into the charging socket of the device the treatment cannot be conducted The...

Page 11: ...atment Since the causes of hyperhidrosis differ between patients you will have to adjust the therapy to your individual needs and lifestyle These factors may vary according to yearly seasons and every...

Page 12: ...vel surface For the treatment of the hands place both trays side by side on a table for the treatment of the feet place both trays side by side on the floor If you treat hands and feet simultaneously...

Page 13: ...your hands out of the tub at any time to readjust the amperage Please note that you should only increase the amperage slowly The increase of the current strength is restricted which means that turnin...

Page 14: ...up the amperage until the current is noticeable then turn it down again until you can no longer feel it Pulsed current Turn up the amperage until the current is barely noticeable Iontophoresis treatme...

Page 15: ...on the display for 5 seconds this indicates that the device is running Now check whether you have selected the correct current type The first line on the display shows if the device is set to DC dire...

Page 16: ...hands or feet If necessary adjust the water level The current flow will be activated as soon as the electric circuit is closed The current strength will increase until the desired value has been reac...

Page 17: ...line on the display shows if the device is set to DC direct current or PC pulsed current When turning the button 1 clockwise the amperage will be increased The second line on the display 5 shows the c...

Page 18: ...ts prefer the comfortable treatment feeling of the pulsed current To turn on the device turn the switch 1 to the left A text will appear on the display for 5 seconds this indicates that the device is...

Page 19: ...aused treatment every third day two times a week with a minimum pause of two days The duration is fixed to 20 minutes to allow an evaluation of the effectiveness of the treatment this leads to a reduc...

Page 20: ...ue to this the device can be operated independently from a power connection Monitored battery accumulator capacity As soon as the battery voltage drops below a defined value the yellow LED 4 will ligh...

Page 21: ...V peak Notice When the DC plug of the charger is plugged into the charging socket the treatment cannot be conducted The DC plug disconnects the electric circuit of the power supply Before using the de...

Page 22: ...eries out For short term storage please check or renew the batteries periodically to keep the equipment in a fully usable condition Disposal The model DE20 iontophoresis device contains electronic com...

Page 23: ...n Read operator manual Consult instruction for use Manufacturer Manufacture date Batch code IP21 Protected against solid objects over 12 5 mm Protected against vertically falling drops of water 2460 C...

Page 24: ...device is suitable for use in all establish ments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domesti...

Page 25: ...floors are covered with synthetic material the relative humidity should be at least 30 Power frequency 50 or 60 Hz magnetic field IEC 61000 4 8 30 A m 50 or 60 Hz 30 A m 50 or 60 Hz Power frequency ma...

Page 26: ...ital environ ment If the user of the device requires continued operation during power mains interruptions it is re commended that the device be powered from an uninterrupti ble power supply or a batte...

Page 27: ...mitter in watts W according to the trans mitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF trans mitters as determined by an electromagnetic si...

Page 28: ...Cosmetics Company AG Nauenstr 67 4052 Basel Switzerland Distribution Center EU Schwarzwaldstr 31 79539 L rrach Germany Tel 41 61 262 10 00 Fax 41 61 303 80 09 contact functional cosmetics com www swea...

Reviews: