28
I n s t r u c t i o n M a n u a l M o d e l D R X - 4 C
Note:
For more on the counting feature, please turn to Chapters 12 and
13 of this manual.
23
6. Once the new sample size has been selected, press the T button to
confirm the new sample size selection. The display will now read SAPL On,
indicating you should to place a new sample of 10 pieces on the pan (if SPL-
10 is selected).
Note:
Minimum individual piece weight allowed is 30mg (3e). With a
sample size of 10 pieces the sample weight must be equal to
or greater than 300mg (30e). Attempting to set a sample
size of 10 pieces with a total sample weight which is lower
than 300mg (30e) will result in a Sample Low error (SAPL-LO)
7. The weight of the new sample will be displayed as it is placed on the pan.
8. Once the weight of the new sample stabilizes and the stable indicator
appears on the display, press the T button to confirm and accept the new
sample weight.
9. The scale will now store the new sample weight and sample size in the
internal memory. The display will indicate you should begin filling, and the
counting transaction may be continued based on the new sample size and
sample weight.