47
14. Specification
■
Specification
CAUTION
For quality improvement, specifications are subject to change without prior
notice.
Standard Specification
Size:
60.6(W) x 60.0(H) x 20.6(D)mm (not including the
protrusion)
Weight:
Approximately 85 grams (with battery)
Transmitting
Waveform:
ECG 1CH or 2CH (selectable from the ECG lead
cable), Respiration waveform
ECG Lead cable Type: 3-electrode, 4-electrode, or 5-electrode (Limb+Chest)
lead cable
Automatically detect the type after inserting the lead
cable
Transmitting Status
Data:
Electrode Off, Low Battery, Event Switch, Pacemaker
Detection, Channel ID, 64group Codes
LCD: Built-in
Waterproof: IPX8
Power Supply:
DC: one (1) 1.5 V “AA” size (“LR06” size) alkaline
battery
Continuous Operating
Time:
Approximately 6 days
※
Continuous operating time is assumed when using new “AA” size (“LR06”
size) Alkaline battery specified by Fukuda Denshi.
ECG
Numbers of Lead
Electrode:
3-electrode, 4-electrode, or 5-electrode
(Limb+Chest)
Numbers of Input
Channel:
1CH(3-electrode) or 2CH
Accuracy of Sensitivity 10mm /1mV ±1mm
(Display sensitivity on the receiving monitor)
ECG Input Impedance: 5M
Ω
and above
Maximum Input Voltage: ±5mV and above
Common Mode Rejection
Ratio:
Less than 10mVp-p (95dB and above)
Accuracy of Heart Rate
Measurement:
±10% or ±5bpm, whichever is greater
Display Coverage of
Heart Rate:
0, 12 to 300bpm (1bpm step)
Frequency
Characteristic:
0.5 to 40Hz (within -3dB)
Summary of Contents for LX-7120
Page 18: ...xvi Blank Page ...
Page 22: ...1 General Description 2 Blank Page ...
Page 24: ...2 Names of Parts and Their Functions 4 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 44: ...6 Measurement 24 Blank Page ...
Page 50: ...7 Operation 30 Blank Page ...
Page 54: ...8 Other Setting Items 34 Blank Page ...
Page 56: ...9 Changing the Transmitter Channel and Group ID 36 Blank Page ...
Page 62: ...11 Cleaning and Disinfection 42 Blank Page ...
Page 73: ......