viii
Precautions about the Maintenance
Safety Inspection and Maintenance
For safe operation of the equipment, regular inspection and maintenance is required. Once a year,
check all cables, devices, and accessories for damage, earth impedance, earth and enclosure leakage
currents, and all alarm functions. Also, ensure that all safety labels are legible. Maintain a record of
these safety inspections.
Immediate maintenance has to be carried out if ;
?
the equipment was subjected to extreme mechanical stress, e.g. after a heavy fall.
?
the equipment was subjected to liquid spill.
?
the monitoring function is interrupted or disturbed.
?
parts of the equipment enclosure are cracked, removed, or lost.
?
any connector or cable shows signs of deterioration.
Reference
Refer to “10. Maintenance” for details.
W A R N I N G
Never open the housing while the equipment is in operation or connected to
hospital grade outlet as it may result in electric shock.
Maintenance, Modifications, and Repairs
Fukuda Denshi is liable for the safety, reliability, and performance of its equipment only if;
?
Maintenance, modifications, and repairs are carried out by authorized personnel.
?
Components are used in accordance with Fukuda Denshi operating instructions.
A full technical description of the DS-7100 system is available from your local Fukuda Denshi
representative.